8 Routine Maintenance<br />
Warning<br />
On a DAILY basis, inspect the patient circuit to ensure that it shows no signs of damage, is<br />
properly connected, and is operating correctly without leakage.<br />
Do not attempt to open, repair or otherwise service the ventilator yourself. Doing so might<br />
endanger the patient, damage the ventilator, and/or void your warranty. Only personnel<br />
authorised and qualified by <strong>Covidien</strong> should repair, open or service the ventilator.<br />
8.1 Replacing the Air Inlet Filter<br />
Warning<br />
Regularly check the cleanliness of the air inlet filter located on the rear of the ventilator. If<br />
necessary, replace the filter before the recommended replacement period is over. This is<br />
particularly important when the ventilator is installed on a wheelchair, because environmental<br />
conditions may cause the filter to become dirty more rapidly.<br />
Failing to replace a dirty air inlet filter, or operating the ventilator without a filter, may cause<br />
serious damage to the ventilator.<br />
The air inlet filter is not reusable; do not attempt to wash, clean, or reuse it.<br />
If the ventilator is used indoors, the condition of the air inlet filter should be checked monthly. If the ventilator<br />
is used outdoors or in a dusty environment, the air inlet filter should be checked weekly and replaced as necessary.<br />
To replace the air inlet filter:<br />
1. Hold the filter between your fingers (see Figure 8-1, item 1).<br />
2. Remove the filter (Figure 8-1, item 2) and discard it.<br />
3. Place the new filter in the device, while ensuring that:<br />
a. The fine particle side of the filter faces outwards, away from the ventilator.