Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
Puritan Bennett 560 Ventilator User's Manual - Covidien
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
5.3 Starting Ventilation<br />
Operating Procedures<br />
Before starting ventilation, refer to Appendix C, “Operational Verification Checklist”, and set the parameter<br />
values in the Preferences menu.)<br />
Warning<br />
Verify the functionality of the alarm conditions before connecting the patient to the ventilator.<br />
Before starting ventilation, ensure that the device is properly assembled and that the air inlet,<br />
cooling vents, and alarm sound diffusion holes are not obstructed. Ensure also that the patient<br />
circuit is of the proper configuration (double or single limb), properly connected to the<br />
ventilator, and that the circuit hoses are neither damaged nor compressed and contain no<br />
obstructions or foreign bodies.<br />
Note:<br />
Many of the functions of the ventilator are not accessible when the Locking key is enabled. For<br />
additional assistance contact your Clinician or equipment representative.<br />
When the ventilator is in standby (the ventilator is on, but ventilation has not started), a message that prompts<br />
the ventilator operator to press VENTILATION ON/OFF to start ventilation is displayed in the right-hand<br />
window of the ventilation and alarm menus (Figure 5-7).<br />
To start ventilation:<br />
Figure 5-7. Prompt to Start Ventilation<br />
Press and release VENTILATION ON/OFF (Figure 5-8, item 1).<br />
• The blue light indicator, at the upper right of the VENTILATION ON/OFF key<br />
(see Figure 5-8, item 2), turns off.<br />
• A “beep“ sounds.<br />
• The ventilation starts.<br />
• The values of the monitored parameters are displayed in the right-hand window.<br />
User’s <strong>Manual</strong> 5-7