Specifications<br />

Table A-3. Internal Lithium Ion Battery<br />

Voltage 25.2 VDC<br />

Full-load capacity 4.8 Ah<br />

Ampere-hour rating On standby: 1.5 Ah<br />

During ventilation: 0.5 Ah<br />

Watt hour rating<br />

Charging current<br />

124Wh to 126Wh<br />

• Standby mode<br />

1.5 A/hr. (duration: < 6 hr.)<br />

• Ventilation mode<br />

0.5 A/hr. (duration: < 13 hr.)<br />

Average operating time at 25 °C (± 5 °C) with a fully charged battery (having less than<br />

50 charge/discharge cycles) at the following displayed values:<br />

Vt = 200 ml (± 5 ml), PIP = 10 mbar (± 2 mbar), Rtot = 20 bpm 11 hr. (–10%)<br />

Vt = 300 ml (± 5 ml), PIP = 20 mbar (± 2 mbar), Rtot = 15 bpm 9 hr. (–10%)<br />

Vt = 500 ml (± 5 ml), PIP = 30 mbar (± 2 mbar), Rtot = 15 bpm 6.5 hr. (–10%)<br />

Vt = 750 ml (± 5 ml), PIP = 45 mbar (± 2 mbar), Rtot = 20 bpm<br />

(maximum settings)<br />

4.5 hr. (–10%)<br />

Table A-4. Remote Alarm<br />

Remote Alarm Port:<br />

Also known as the Nurse’s Call port, it<br />

provides for remote alerts of ventilator<br />

alarm conditions.<br />

An example of a setting that requires such<br />

a feature is when the ventilator is used in<br />

an isolation room.<br />

The ventilator signals an alarm using a<br />

normally open (NO) or a normally closed<br />

(NC) signal.<br />

A remote alarm is activated when an alarm<br />

condition occurs, unless either of the<br />

following is true:<br />

• Audio paused function is active<br />

• <strong>Ventilator</strong> power switch is turned off<br />

The remote alarm port is an 8-pin female<br />

connector. Allowable current is 100 mA at<br />

24 VDC (maximum).<br />

A-2 User’s <strong>Manual</strong><br />

3<br />

2<br />

1<br />

4<br />

5<br />

6<br />

7<br />

8<br />

Nurse call pin-out (view from back of ventilator)<br />

Pin Signal<br />

Remote<br />

Alarm Wire<br />

color<br />

1 relay common black<br />

2 normally open (NO) brown<br />

3 normally closed (NC) orange<br />

4 remote supply - (not<br />

used)<br />

5 RX Signal (not used)<br />

6 TX Signal (not used)<br />

7 remote supply + (not<br />

used)