Operating Procedures<br />
• The parameter selected to be modified flashes.<br />
3. Press UP or DOWN to change the selected parameter’s value.<br />
4. Press ENTER to confirm the new parameter setting.<br />
• The new parameter setting is displayed continuously.<br />
• The cursor is placed at the STOP position.<br />
5. To manually stop trend transfer, press ENTER .<br />
If a parameter change is not confirmed by pressing ENTER before seven (7) seconds elapse, the ventilator<br />
resets the parameter to its previous value.<br />
Note:<br />
Table 5-2. Time taken to transfer trends data from the ventilator to a USB memory device<br />
Amount of trends data<br />
(in months)<br />
Transfer time from<br />
ventilator to USB memory<br />
device<br />
3 months Approximately 2 minutes<br />
6 months Approximately 4 minutes<br />
9 months Approximately 6 minutes<br />
12 months Approximately 8 minutes<br />
The message “TRANSFER IN PROGRESS... REMAINING TIME” is displayed during the transfer time.<br />
Other USB memory device functions are available during transfer of trends.<br />
If the memory capacity on the USB memory device is insufficient the message “TRANSFER NOT ALLOWED -<br />
USB CAPACITY INSUFFICIENT” is displayed and data transfer is not allowed. Delete the data on the USB<br />
memory device before restarting data transfer.<br />
In case of USB memory device disconnection or transfer error, the message “TRANSFER ERROR - USB<br />
DISCONNECTION” or “TRANSFER ERROR - TECHNICAL PROBLEM” is displayed. In this case restart the<br />
transfer process. If the problem persists contact your technical service.<br />
5-6 User’s <strong>Manual</strong>