10.10.2013 Views

Protein Expression and Purification Series - Bio-Rad

Protein Expression and Purification Series - Bio-Rad

Protein Expression and Purification Series - Bio-Rad

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Protein</strong> <strong>Expression</strong> <strong>and</strong> <strong>Purification</strong> <strong>Series</strong><br />

Final Check Before Starting Method<br />

1.<br />

2.<br />

3.<br />

4.<br />

Check for leaks or air in the lines.<br />

Check to make sure that your buffer inlet lines are at the bottom of the beakers containing Buffer A<br />

<strong>and</strong> Buffer B.<br />

Make sure that you have at least 50 ml of Buffer A <strong>and</strong> 20 ml of Buffer B.<br />

Place collection tubes in the fraction collector rack in the first two columns (30 tubes total) for the<br />

<strong>Bio</strong>Frac fraction collector or the first 30 spaces for the Model 2110 fraction collector.<br />

5. If you are using LP DataView software, press STOP on the software menu. Press CLEAR to clear<br />

all previous information. A menu box will appear. Choose “NO.” Press RECORD to record your<br />

purification run.<br />

Load Sample Loop with GST-DHFR-His Soluble Fraction<br />

1.<br />

2.<br />

3.<br />

4.<br />

5.<br />

Make sure the MV-6 injector valve knob counter is turned<br />

counterclockwise (left) as far as it will go (Figure 7.10).<br />

Draw 5 ml of the soluble fraction of GST-DHFR-His into the 10 ml<br />

syringe. Avoid pulling in any air bubbles. If you have more than 5 ml of<br />

soluble fraction, you can draw the entire fraction into the syringe.<br />

Make sure to have the waste port tubing of the MV-6 injector valve<br />

over a waste beaker.<br />

Insert the syringe in the top port of the MV-6 injector valve <strong>and</strong> fill the<br />

sample loop. The excess water that was in the sample loop from the<br />

cleaning of the loop will be pushed out of the waste port.<br />

Note: Avoid pushing air bubbles into the sample loop since these will<br />

flow into the column <strong>and</strong> through the detectors.<br />

Leave the syringe in the port when you have injected the sample; this prevents the sample from<br />

siphoning out of the loop.<br />

Start Method<br />

1. Press the Run mode key.<br />

2. Immediately turn the injector valve knob clockwise<br />

(to the right) as far as it will go (Figure 7.11).<br />

Figure 7.11. MV-6 injection port in the INJECT position. The<br />

inject position (turned fully clockwise) opens up the fluid pathway<br />

so that the buffer coming from the pump will push the sample out<br />

of the sample loop <strong>and</strong> onto the column.<br />

Chapter 7: <strong>Purification</strong> Protocol for <strong>Bio</strong>Logic LP System<br />

149<br />

Figure 7.10. MV-6 in Sample<br />

Load position, turned as far left<br />

as possible.<br />

CHAPTER 7<br />

BIOLOGIC LP SYSTEM<br />

PROTOCOL

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!