Appendix E: Gel Box Setup<br />

The arms of the clamping frame push the short plates of each gel cassette up against the notch<br />

in the green gasket, creating a leak-proof seal (check again to make certain that the short plates<br />

sit just below the notch at the top of the green gasket). At this point, the sample wells can be<br />

washed-out with running buffer (Figure 2D).<br />

Note: If running more than two gels, repeat steps 1A–D with the Companion Running Module.<br />

Important Note: Do not attempt to lock the green arms of the clamping frame, without first<br />

ensuring that the gel cassettes are perfectly aligned <strong>and</strong> stabilized against the notches on the<br />

green gaskets of the module. To prevent the gels from shifting during the locking step, firmly <strong>and</strong><br />

evenly grip them in place against the core of the module with one h<strong>and</strong>.<br />

Run the gel(s): Prepare, load <strong>and</strong> electrophorese the samples according to the instructions<br />

in Chapters 4 or 5, depending on whether you are performing the centrifugation process or<br />

chromatography instrumentation process for purification. At the completion of the run, disconnect<br />

the cell <strong>and</strong> remove the cassette.<br />

CAUTION: When running one or two gels only, DO NOT place the Companion Running<br />

Module in the tank. Doing so will cause excessive heat generation <strong>and</strong> prevent<br />

electrophoretic separation.<br />

Removing the Gel from the Cassette after the Gel has Been Run<br />

1. Open the cassette:<br />

Align the arrow on the opening lever with the arrows marked on the<br />

cassette (see Figure 3). Insert the lever between the cassette plates at all four locations <strong>and</strong> apply<br />

downward pressure to break each seal. Do not twist the lever. Gently pull apart the two plates<br />

beginning from the top of the cassette.<br />

2. Remove Gel:<br />

Gently remove the gel from the cassette.<br />

Figure 3. Opening the gel cassette.<br />

225<br />

APPENDIX E<br />

MINI-PROTEAN GEL<br />