CHAPTER 8<br />

BIOLOGIC DUOFLOW<br />

PROTOCOL<br />

<strong>Protein</strong> <strong>Expression</strong> <strong>and</strong> <strong>Purification</strong> <strong>Series</strong><br />

Chapter 8: <strong>Purification</strong> Protocol for <strong>Bio</strong>Logic DuoFlow<br />

System<br />

Affinity <strong>Purification</strong> of GST-DHFR-His Using a <strong>Bio</strong>Logic<br />

DuoFlow System<br />

Student Workstations<br />

Each student team requires the following items to purify their GST-DHFR-His samples using affinity<br />

chromatography:<br />

Material Needed for Each Workstation Quantity<br />

Soluble lysate fraction 5–7.5 ml<br />

10 ml syringe 1<br />

Distilled water 500 ml<br />

Buffer A 70 ml<br />

Buffer B 40 ml<br />

Collection tubes for fraction collector 32<br />

Marking pen 1<br />

Common Workstation Quantity<br />

<strong>Bio</strong>Logic DuoFlow chromatography instrument<br />

with 5 ml sample loop <strong>and</strong> column installed:<br />

Either a <strong>Bio</strong>Scale Mini Profinity Ni-IMAC Cartridge<br />

or<br />

1–4<br />

A glass Econo-Column column h<strong>and</strong>packed with 1 ml of Profinity IMAC Ni-charged resin<br />

<strong>Bio</strong>Logic DuoFlow injection needle 1–4<br />

Preparing the DuoFlow system for Use<br />

1. Turn on the power switches for all hardware (e.g. Maximizer, if you have one), Workstation, <strong>and</strong><br />

<strong>Bio</strong>Frac fraction collector. Turn on the power for the Controller (computer) once all the other hardware<br />

has been turned on.<br />

2.<br />

3.<br />

Double click on the <strong>Bio</strong>Logic Configuration icon.<br />

Make sure that the configuration is set up for the system that you have installed (Figure 8.1), then click<br />

“OK” to close out the Configuration Utility.<br />

4. Double click on the <strong>Bio</strong>Logic DuoFlow application icon to launch the software.<br />