22.07.2013 Views

Principles of Fluorescence Spectroscopy

Principles of Fluorescence Spectroscopy

Principles of Fluorescence Spectroscopy

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

PRINCIPLES OF FLUORESCENCE SPECTROSCOPY 177<br />

Figure 5.25. Frequency response <strong>of</strong> 4-dimethylamino-4'-bromostilbene<br />

(DBS) up to 10 GHz. The vertical dashed lines are at 200 MHz,<br />

2 GHz, and 10 GHz. From [3].<br />

Figure 5.24 shows that the photodiode provides a higher<br />

bandwidth than does any <strong>of</strong> the MCP PMTs. In fact, photodiode<br />

detectors were used in several phase fluorometers<br />

for measurements at high frequencies. 109–112 Unfortunately,<br />

the small active area <strong>of</strong> photodiodes results in low sensitivity,<br />

so that photodiodes are rarely used for fluorescence<br />

spectroscopy.<br />

The schematic for the 10-GHz instrument shown in<br />

Figure 5.23 incorporates a ps laser as an intrinsically modulated<br />

light source, and an MCP PMT as the detector. A<br />

photodiode (PD) is adequate as the reference detector<br />

because the laser beam can be focused on its small active<br />

area. The use <strong>of</strong> cross-correlation allows measurement over<br />

the entire frequency range from 1 MHz to 10 GHz without<br />

any noticeable increase in noise. Cross-correlation allows<br />

measurements at any modulation frequency at the same low<br />

cross-correlation frequency, and avoids the need to measure<br />

phase angles and modulation at high frequencies. A valuable<br />

feature <strong>of</strong> cross-correlation is that the entire signal<br />

appears at the measured frequency. Contrary to intuition,<br />

one is not selecting one harmonic component out <strong>of</strong> many,<br />

which would result in low signal levels. The use <strong>of</strong> crosscorrelation<br />

provides absolute phase and modulation values<br />

as if the excitation and detector were both modulated as<br />

sine waves. A final favorable feature <strong>of</strong> this instrument is<br />

that the modulation can be higher than 1.0, which is the<br />

limit for sine wave modulation. At low frequencies where<br />

the detector is fully responsive, the modulation can be as<br />

high as 2.0. To understand this unusual result one needs to<br />

examine the Fourier components for a pulse train.<br />

5.6.1. Gigahertz FD Measurements<br />

Several examples <strong>of</strong> gigahertz FD measurements will illustrate<br />

the value <strong>of</strong> a wide range <strong>of</strong> frequencies. Short decay<br />

times are needed to utilize the high-frequency capabilities.<br />

113 Otherwise, the emission is demodulated prior to<br />

reaching the upper frequency limit. A short decay time was<br />

obtained using 4-dimethylamino-4'-bromostilbene (DBS)<br />

in cyclohexane at 75EC (Figure 5.25). Because <strong>of</strong> the short<br />

61-ps lifetime the phase and modulation data could be<br />

measured to 10 GHz. The intensity decay was found to be a<br />

single exponential. 3 The vertical dashed lines illustrate how<br />

only a fraction <strong>of</strong> the frequency response could be explored<br />

if the upper limit was 200 MHz, or even 2 GHz. It would be<br />

difficult to detect additional components in the intensity<br />

decay if the data stopped at 200 MHz, which would display<br />

a maximum phase angle <strong>of</strong> 4.4E. An important aspect <strong>of</strong><br />

these measurements is that no measurable color effect has<br />

been observed in the 10 GHz measurements. 106<br />

5.6.2. Biochemical Examples <strong>of</strong><br />

Gigahertz FD Data<br />

GHz measurements may seem exotic, but such data are<br />

<strong>of</strong>ten needed for studies <strong>of</strong> routine biochemical samples.<br />

One example is NADH. At 200 MHz the data only sampled<br />

part <strong>of</strong> the frequency response (Figure 5.16). When measured<br />

to higher frequencies one can more dramatically see<br />

the difference between the one and two decay time fits (Figure<br />

5.26). The decrease in χ R 2 for the two decay time model<br />

is 400-fold. While we have not performed a support plane<br />

analysis on these data, the α i and τ i values will be more<br />

closely determined using the data extending above 200<br />

MHz.<br />

Another biochemical example is provided by the peptide<br />

hormone vasopressin, which acts as an antidiuretic and

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!