PRINCIPLES OF FLUORESCENCE SPECTROSCOPY 369<br />

Figure 10.14. Perrin plots for determination <strong>of</strong> protein volume. The<br />

curves are drawn for a protein molecular weight <strong>of</strong> 25,000 daltons in<br />

water, h = 0.4 and τ = 4 ns. The value <strong>of</strong> r 0 = 0.4 represents a rigidly<br />

bound probe and r 0 = 0.3 represents a probe with some segmented<br />

mobility. The upward curving dashed lines indicate the protein is<br />

denaturing at high temperature.<br />

For example, assume the segmental motion <strong>of</strong> the fluorophore<br />

is much more rapid than the rotational diffusion <strong>of</strong><br />

the protein. Then, the value <strong>of</strong> r 0 is effectively reduced to<br />

r app<br />

0 r0 ( 1<br />

2<br />

(10.48)<br />

where α is the angle through which the probe undergoes<br />

this segmental motion. The term (1/r 0 app) cancels in the calculation<br />

<strong>of</strong> the molecular volume, and the apparent volume<br />

represents overall rotation diffusion <strong>of</strong> the protein. It is<br />

important to note that, if the short correlation time for segmental<br />

motion (θ S ) is not much faster than the overall correlation<br />

time (θ L ), then the apparent volume can be substantially<br />

decreased due to contributions from the more rapid<br />

motions (Section 10.9). Some information is available by<br />

comparison <strong>of</strong> the extrapolated and frozen solution values<br />

<strong>of</strong> r 0 . If r 0 app is much smaller than r 0 one may infer the existence<br />

<strong>of</strong> segmental motions <strong>of</strong> the probe on the macromol-<br />

)<br />

ecule, independent motions <strong>of</strong> domains <strong>of</strong> the protein, or<br />

possibly RET in multiply labeled proteins. Another behavior<br />

seen in Perrin plots is a rapid increase in 1/r at high temperatures.<br />

Such an increase is usually the result <strong>of</strong> denaturation<br />

<strong>of</strong> the macromolecule, resulting in an increase in<br />

independent motion <strong>of</strong> the probe and a decrease in the<br />

anisotropy.<br />

A different version <strong>of</strong> the Perrin equation is <strong>of</strong>ten<br />

found in the older literature:<br />

( 1<br />

P<br />

1 1<br />

) ( 3 P0 1 τRT<br />

) ( 1 <br />

3 ηV )<br />

( 1<br />

<br />

P0 1<br />

3 ) ( 1 3τ<br />

ρ )<br />

(10.49)<br />

This equation is equivalent to eq. 10.47, except for the use<br />

<strong>of</strong> polarization in place <strong>of</strong> anisotropy. When using the equation,<br />

a plot (1/P – 1/3) versus T/η is used to obtain the<br />

molecular volume. The intercept yields the apparent value<br />

<strong>of</strong> (1/P 0 app – 1/3), which can be larger than the true value<br />

(1/P 0 – 1/3) if there is segmental motion <strong>of</strong> the probe. The<br />

term ρ is the rotational relaxation time (ρ = 3θ). At present<br />

the use <strong>of</strong> anisotropy, the rotational correlation time, and eq.<br />

10.44 is preferred.<br />

10.5.2. Examples <strong>of</strong> a Perrin Plot<br />

It is instructive to examine a representative Perrin plot (Figure<br />

10.15). 16 These data are for 9-anthroyloxy stearic acid<br />

in a viscous paraffin oil, Primol 342. The anisotropies were<br />

measured at various excitation wavelengths, corresponding<br />

to different r 0 values (Figure 10.8). The y-axis intercepts are<br />

different because <strong>of</strong> the different r 0 values. Also, the slopes<br />

are larger for shorter excitation wavelengths because the<br />

value <strong>of</strong> r 0 appears in the denominator <strong>of</strong> eq. 10.47. In spite<br />

<strong>of</strong> the very different values <strong>of</strong> 1/r for different excitation<br />

wavelengths, the data are all consistent with a correlation<br />

time near 15 ns at 25EC (Table 10.5). For a spherical molecule,<br />

the correlation time is independent <strong>of</strong> r 0 . In Chapter 12<br />

we will see that the correlation times can depend on the<br />

value <strong>of</strong> r 0 for non-spherical molecules. This effect may be<br />

the cause <strong>of</strong> the somewhat different correlation times found<br />

for different excitation wavelengths.<br />

The y-axis intercepts yield the apparent values <strong>of</strong> r 0<br />

extrapolated to high viscosity, which can be compared with<br />

measured values in a glassy solvent (Table 10.5). The