Lumenera <strong>INFINITY</strong> <strong>ANALYZE</strong><br />
User’s <strong>Manual</strong> Release 6.0.0<br />
3 From Capture Options Sub-Pane<br />
Select the number of frames to be averaged. If 1 is<br />
selected, no frame averaging will be performed. If<br />
2 is selected, two successive frames will be<br />
summed and then divided by two to create the<br />
intermediate image that is the input to subsequent<br />
post-capture processing. 4 and 8 are similarly<br />
defined.<br />
Frame averaging reduces noise. If n frames are<br />
averaged, the energy of the noise will be 1/n of its<br />
original level.<br />
Frame averaging is performed only when capturing<br />
full field of view.<br />
Sub-sample the output image. If 1 is selected, no<br />
sub-sampling will be done. If 2 is selected, the<br />
number of columns and number of rows of the<br />
image will be reduced by a factor of two by<br />
merging every two consecutive columns and rows.<br />
3 and 4 are similarly defined.<br />
The way <strong>INFINITY</strong> <strong>ANALYZE</strong> performs<br />
Page 42 Copyright © 2006-2011