Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
170 Chapter 6<br />
<strong>to</strong> be analyzed, whereas the XWDS analyzing crystal needs room <strong>to</strong> move<br />
and subtends a smaller (solid) angle, so that much less than 1% <strong>of</strong> the x-ray<br />
pho<strong>to</strong>ns are collected. Although a commercial XEDS system costs $50,000<br />
or more, this is considerably less expensive than an EPMA machine or<br />
XWDS<br />
attachment <strong>to</strong> an SEM.<br />
The main advantage <strong>of</strong> the XWDS system is that it provides narrow xray<br />
peaks (width � 5 eV, rather than > 100 eV for XEDS system) that stand<br />
out clearly from the background; see Fig. 6-7. This is particularly important<br />
when analyzing low-Z elements (whose K-peaks are more narrowly spaced)<br />
or elements present at low concentration, whose XEDS peaks would have a<br />
low signal/background ratio, resulting in a poor accuracy <strong>of</strong> measurement.<br />
Consequently, XWDS analysis is the preferred method for measuring low<br />
concentrations: 200 parts per million (ppm) is fairly routine and 1 ppm is<br />
detectable in special cases. At higher concentrations, the narrow XWDS<br />
peak allows a measurement accuracy <strong>of</strong> better than 1% with the aid <strong>of</strong> ZAF<br />
corrections. In contrast, the accuracy <strong>of</strong> XEDS analysis is usually not better<br />
than 10%.<br />
Figure 6-7. X-ray intensity as a function <strong>of</strong> pho<strong>to</strong>n energy (in the range 2.1 <strong>to</strong> 2.7 keV),<br />
recorded from a molybdenum disulfide (MoS2) specimen using (a) XWDS and (b) XEDS.<br />
The wavelength-dispersed spectrum displays peaks due <strong>to</strong> Mo and S, whereas these peaks are<br />
not separately resolved in the energy-dispersed spectrum due <strong>to</strong> the inadequate energy<br />
resolution Courtesy <strong>of</strong> S. Matveev, <strong>Electron</strong> Microprobe Labora<strong>to</strong>ry, University <strong>of</strong> Alberta.