16.11.2012 Views

Physical Principles of Electron Microscopy: An Introduction to TEM ...

Physical Principles of Electron Microscopy: An Introduction to TEM ...

Physical Principles of Electron Microscopy: An Introduction to TEM ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

170 Chapter 6<br />

<strong>to</strong> be analyzed, whereas the XWDS analyzing crystal needs room <strong>to</strong> move<br />

and subtends a smaller (solid) angle, so that much less than 1% <strong>of</strong> the x-ray<br />

pho<strong>to</strong>ns are collected. Although a commercial XEDS system costs $50,000<br />

or more, this is considerably less expensive than an EPMA machine or<br />

XWDS<br />

attachment <strong>to</strong> an SEM.<br />

The main advantage <strong>of</strong> the XWDS system is that it provides narrow xray<br />

peaks (width � 5 eV, rather than > 100 eV for XEDS system) that stand<br />

out clearly from the background; see Fig. 6-7. This is particularly important<br />

when analyzing low-Z elements (whose K-peaks are more narrowly spaced)<br />

or elements present at low concentration, whose XEDS peaks would have a<br />

low signal/background ratio, resulting in a poor accuracy <strong>of</strong> measurement.<br />

Consequently, XWDS analysis is the preferred method for measuring low<br />

concentrations: 200 parts per million (ppm) is fairly routine and 1 ppm is<br />

detectable in special cases. At higher concentrations, the narrow XWDS<br />

peak allows a measurement accuracy <strong>of</strong> better than 1% with the aid <strong>of</strong> ZAF<br />

corrections. In contrast, the accuracy <strong>of</strong> XEDS analysis is usually not better<br />

than 10%.<br />

Figure 6-7. X-ray intensity as a function <strong>of</strong> pho<strong>to</strong>n energy (in the range 2.1 <strong>to</strong> 2.7 keV),<br />

recorded from a molybdenum disulfide (MoS2) specimen using (a) XWDS and (b) XEDS.<br />

The wavelength-dispersed spectrum displays peaks due <strong>to</strong> Mo and S, whereas these peaks are<br />

not separately resolved in the energy-dispersed spectrum due <strong>to</strong> the inadequate energy<br />

resolution Courtesy <strong>of</strong> S. Matveev, <strong>Electron</strong> Microprobe Labora<strong>to</strong>ry, University <strong>of</strong> Alberta.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!