Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>An</strong> <strong>Introduction</strong> <strong>to</strong> <strong>Microscopy</strong> 21<br />
1.7 <strong>An</strong>alytical <strong>Electron</strong> <strong>Microscopy</strong><br />
All <strong>of</strong> the images seen up <strong>to</strong> now provide information about the structure <strong>of</strong><br />
a specimen, in some cases down <strong>to</strong> the a<strong>to</strong>mic scale. But <strong>of</strong>ten there is a need<br />
for chemical information, such as the local chemical composition. For this<br />
purpose, we require some response from the specimen that is sensitive <strong>to</strong> the<br />
exact a<strong>to</strong>mic number Z <strong>of</strong> the a<strong>to</strong>ms. As Z increases, the nuclear charge<br />
increases, drawing electrons closer <strong>to</strong> the nucleus and changing their energy.<br />
The electrons that are <strong>of</strong> most use <strong>to</strong> us are not the outer (valence) electrons<br />
but rather the inner-shell electrons. Because the latter do not take part in<br />
chemical bonding, their energies are unaffected by the surrounding a<strong>to</strong>ms<br />
and remain indicative <strong>of</strong> the nuclear charge and therefore a<strong>to</strong>mic number.<br />
When an inner-shell electron makes a transition from a higher <strong>to</strong> a lower<br />
energy level, an x-ray pho<strong>to</strong>n is emitted, whose energy (hf = hc/�) is equal <strong>to</strong><br />
the difference in the two quantum levels. This property is employed in an xray<br />
tube, where primary electrons bombard a solid target (the anode) and<br />
excite inner-shell electrons <strong>to</strong> a higher energy. In the de-excitation process,<br />
characteristic x-rays are generated. Similarly, the primary electrons<br />
entering a <strong>TEM</strong>, SEM, or S<strong>TEM</strong> specimen cause x-ray emission, and by<br />
identifying the wavelengths or pho<strong>to</strong>n energies present, we can perform<br />
chemical (more correctly: elemental) analysis. Nowadays, an x-ray emission<br />
spectrometer is a common attachment <strong>to</strong> a <strong>TEM</strong>, SEM, or S<strong>TEM</strong>, making<br />
the instrument in<strong>to</strong> an analytical electron microscope (AEM).<br />
Other forms <strong>of</strong> AEM make use <strong>of</strong> characteristic-energy Auger electrons<br />
emitted from the specimen, or the primary electrons themselves after they<br />
have traversed a thin specimen and lost characteristic amounts <strong>of</strong> energy. We<br />
will examine all <strong>of</strong> these options in Chapter 6.<br />
1.8 Scanning-Probe Microscopes<br />
The raster method <strong>of</strong> image formation is also employed in a scanning-probe<br />
microscope, where a sharply-pointed tip (the probe) is mechanically scanned<br />
in close proximity <strong>to</strong> the surface <strong>of</strong> a specimen in order <strong>to</strong> sense some local<br />
property. The first such device <strong>to</strong> achieve really high spatial resolution was<br />
the scanning tunneling microscope (STM) in which a sharp conducting tip<br />
is brought within about 1 nm <strong>of</strong> the sample and a small potential difference<br />
(� 1 V) is applied. Provided the tip and specimen are electrically conducting,<br />
electrons move between the tip and the specimen by the process <strong>of</strong><br />
quantum-mechanical tunneling. This phenomenon is a direct result <strong>of</strong> the<br />
wavelike characteristics <strong>of</strong> electrons and is analogous <strong>to</strong> the leakage <strong>of</strong>