Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
Physical Principles of Electron Microscopy: An Introduction to TEM ...
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>An</strong>alytical <strong>Electron</strong> <strong>Microscopy</strong> 171<br />
6.8 Auger-<strong>Electron</strong> Spectroscopy<br />
Both XEDS or XWDS techniques have problems when applied <strong>to</strong> the<br />
analysis <strong>of</strong> elements <strong>of</strong> low a<strong>to</strong>mic number (Z < 11) such as B, C, N, O, F.<br />
Ultrathin-window XEDS detec<strong>to</strong>rs make these elements visible, but the<br />
characteristic-peak intensities are reduced because <strong>of</strong> the low x-ray<br />
fluorescence yield, which falls continuously with decreasing a<strong>to</strong>mic number;<br />
see Fig. 6-8. In addition, the K-emission peaks (which must be used for low-<br />
Z elements) occur at energies below 1 keV. In this restricted energy region,<br />
the bremsstrahlung background is relatively high, and XEDS peaks from<br />
different elements tend <strong>to</strong> overlap. Also, the low-energy x-rays are strongly<br />
absorbed, even within a thin specimen, requiring a substantial (and not<br />
always<br />
accurate) correction for absorption.<br />
One solution <strong>to</strong> the fluorescence-yield problem is <strong>to</strong> use Auger electrons<br />
as the characteristic signal. Because the x-ray yield � is low, the Auger yield<br />
(1 ��) is close <strong>to</strong> one for low-Z elements. Auger electrons can be collected<br />
and analyzed using an electrostatic spectrometer, which measures their<br />
kinetic energy. However, the Auger electrons <strong>of</strong> interest have relatively low<br />
energy (below 1000 eV); they are absorbed (through inelastic scattering) if<br />
they are generated more than one or two nm below the surface <strong>of</strong> the<br />
specimen, just as with SE1 electrons. In consequence, the Auger signal<br />
measures a chemical composition <strong>of</strong> the surface <strong>of</strong> a specimen, which can be<br />
1<br />
0<br />
30<br />
x-ray yield<br />
Auger yield<br />
Figure 6-8. X-ray fluorescence yield, Auger yield and their sum (dashed line) for K-electron<br />
excitation, as a function <strong>of</strong> a<strong>to</strong>mic number.<br />
Z