<strong>G4</strong> <strong>Multi</strong> <strong>Channel</strong> <strong>Weighing</strong> <strong>Instrument</strong>(Continued)ErrorcodeExplanation193 Too high transducer signal in calibration point 2The mV/V signal in calibration point 2 is too high, due to strange data sheetcalibration. The conversion factor, rated load etc. does not correspond toeach other.194 Illegal calibration directionAll weights with corresponding mV/V values must be increasing forincreasing calibration point number.195 Illegal I/O module configurationThe I/O-module set-up (Hardware Configuration menu) is not consistentwith the allowed number of modules of the different types or the allowedcombinations of modules.196 Illegal scale number configurationEach scale number must only be used once. I.e. each measuring inputmust have a unique scale number. See menu ‘Hardware Configuration’.7-7