<strong>Technical</strong> <strong>Manual</strong>for the selected number of points.It is possible to change the number of points during the calibration.Value Cal. P1,This parameter defines the load for the lowest calibration point. Normally the scaleshould be unloaded and the parameter value set to 0 (zero). This weight value andthe corresponding transducer signal value are automatically stored in the instrument.Value Cal. P2, Value Cal. P3 etc.The scale should be loaded with known weights. These parameters show the loadaccording to the previous calibration and the parameter values should be changed tothe value of the known weights. As a parameter value is stored, the instrument will alsostore the corresponding transducer signal value for that calibration point.Transd. Signal P1, Transd. Signal P2 etc.These parameters contain the automatically stored transducer signal values forthe calibration points. The values cannot be edited.Set ZeroSet zero is useful only when the scale installation is finished. By the digit keys thisvalue can be set to zero, for unloaded scale, or to the weight of the known load, for ascale that is loaded.Zero OffsetFor an installed scale this parameter shows the zero offset after zeroing, a value thatshould not be edited.4-6