<strong>Technical</strong> <strong>Manual</strong>AOUT1 and AOUT4!The voltage levels on connectors of I/O modules shall not exceedhazardous voltage levels of 30 Vrms, 42.4 Vpeak or 60 Vdc under normalconditions. In wet locations these voltage levels shall not exceed 16 Vrms,22.6 Vpeak or 35 Vdc.Analog output unitThe AOUT4 unit has 4 analog output channels, independentlyisolated by operational insulation.The AOUT1 has 1 output channel, isolated by operationalinsulation.The analog output signal will be connected to terminals10, 11 (channel 1)7, 8 (channel 2)4, 5 (channel 3)1, 2 (channel 4).Shielded cable/cables must be used and the shield(s) must beconnected to the shield terminal(s) 12, 9, 6 and 3.The shields can be connected to the most convenient ground/earthpointAOUT1 <strong>Channel</strong> 1+101112OUT0VShieldAOUT4 <strong>Channel</strong> 4+123OUT0VShield<strong>Channel</strong> 3+456OUT0VShield<strong>Channel</strong> 2+789OUT0VShield<strong>Channel</strong> 1+101112OUT0VShield2-10