30.07.2015 Views

Manual of basic techniques for a health laboratory - libdoc.who.int

Manual of basic techniques for a health laboratory - libdoc.who.int

Manual of basic techniques for a health laboratory - libdoc.who.int

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

278 <strong>Manual</strong> <strong>of</strong> <strong>basic</strong> <strong>techniques</strong> <strong>for</strong> a <strong>health</strong> <strong>laboratory</strong>Fig. 9.26 Calibration carve <strong>for</strong> determining haemoglobin concentration2. Arrange the test-tubes in a test-tube rack: one <strong>for</strong> each sample to be tested, one<strong>for</strong> the blank and two <strong>for</strong> the control samples.3. Using a grease pencil, label the test-tubes with the appropriate <strong>laboratory</strong> numbers<strong>of</strong> the samples to be measured, B <strong>for</strong> the blank, and C1 and C2 <strong>for</strong> thecontrol samples.4. Pipette 3ml <strong>of</strong> AHD reagent <strong>int</strong>o each test-tube.5. Pipette 20ml <strong>of</strong> blood collected in EDTA from a patient <strong>int</strong>o the AHD reagent<strong>of</strong> the appropriate tube. Flush the pipette carefully five times with the AHDreagent.6. Pipette 20ml <strong>of</strong> AHD standard <strong>int</strong>o test-tubes C1 and C2.7. Plug all the test-tubes with a clean cork or rubber stopper and mix by inversion.Leave the tubes to stand <strong>for</strong> 2–3 minutes.8. Pour the AHD solution from test-tube B <strong>int</strong>o a clean cuvette. Dry the outside <strong>of</strong>the cuvette with cotton wool or gauze. Make sure that there are no air bubbles <strong>int</strong>he solution. Place the cuvette in the cuvette chamber and adjust the spectrophotometeror haemoglobinometer to read zero.9. Repeat the procedure with the solution in test-tubes C1 and C2, respectively. Ifthe readings <strong>of</strong> the two controls differ by less than 2.5%, measure the haemoglobinconcentration <strong>of</strong> all the test samples. Record all the results.Method using a colorimeterThe AHD method is also applicable using a colorimeter. The measurement procedureis the same as that described <strong>for</strong> a spectrophotometer or haemoglobinometer.However, the absorbance in a colorimeter does not increase linearly with haemoglobinat elevated concentrations. There<strong>for</strong>e, a calibration curve must be used torelate the absorbance readings to the haemoglobin concentration, as described above.ResultsReport the results in g/l. Example: “haemoglobin = 89g/l”.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!