Net F/T Installation and Operation Manual<br />

Document: 9620-05-net ft-11<br />

18.3 Questions and Answers<br />

18.3.1 Powering Up<br />

Question/Problem<br />

Xdcr LED stays red after the twenty second power up<br />

phase<br />

Xdcr LED is red for the first twenty seconds after<br />

power up then turns green<br />

The LS EN (Ethernet Link Status) is not green or<br />

flashing green<br />

Answer/Solution<br />

Check transducer cable connections.<br />

Verify transducer cable is not damaged.<br />

There may be an internal error in the Net Box.<br />

This is normal operation.<br />

Check Ethernet cable connections.<br />

18.3.2 Communications<br />

Question/Problem<br />

What IP address is assigned to the Net F/T?<br />

How can the Net F/T system be set to the default IP<br />

address of 192.168.1.1?<br />

DHCP is not assigning an IP address<br />

Browser cannot find the Net F/T on Ethernet network<br />

even though the Net F/T configuration utility reports<br />

an IP address<br />

Incorrect CAN Bus Base Address, DeviceNet MAC<br />

ID, and/or Baud Rate reported<br />

<strong>System</strong> status reports DeviceNet Protocol Failure<br />

when using DeviceNet<br />

Answer/Solution<br />

See Section 6.1—Finding Net F/Ts on the <strong>Network</strong>.<br />

Set DIP switch 9 to the ON position (see Section 3.10—DIP<br />

Switches and Termination Resistor). The Net F/T must be power<br />

cycled for the new setting to be used.<br />

Ethernet LAN must be connected during power up.<br />

DHCP is not selected as the IP Address Mode on the<br />

Communications page.<br />

The DHCP server waits more than thirty seconds to respond.<br />

Clear the Windows computer’s ARP table to remove memory of a<br />

previous device that used the same IP address as the Net F/T by<br />

restarting the computer or, if you have administrative privileges, by<br />

going to the computer’s Start menu, selecting Run..., and entering<br />

“arp –d *”.<br />

Power must be present on the Pwr/CAN connector to correctly<br />

report these values.<br />

DeviceNet is not available unless power is present on the Pwr/CAN<br />

connector.<br />

18.3.3 Java Demo<br />

Question/Problem<br />

Demo displays zeros for force and torque values and<br />

question marks for configuration data<br />

Excessive IO exception: Receive timed out errors<br />

Error message: IO exception: <br />

(The process cannot access the file because it is<br />

being used by another process)<br />

The message Could not find the main class. Program<br />

will exit appears in a window titled Java Virtual<br />

Machine Launcher.<br />

Answer/Solution<br />

Check IP address and restart demo.<br />

The Ethernet connection was interrupted. Check Ethernet cabling<br />

and Net F/T power.<br />

Selected file for data is in use by another program. Close file or<br />

change file name and press Collect Streaming again.<br />

Computer requires a newer version of Java. Java may be<br />

downloaded from www.java.com/getjava.<br />

Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 USA • Tel: +1.919.772.0115 • Fax: +1.919.772.8259 • www.ati-ia.com • Email: info@ati-ia.com<br />