Rosemount Raptor<br />

System Configuration Manual<br />

300510EN, Rev AA<br />

December 2010<br />

11. For still pipe applications select the Enable still pipe/Bridle<br />

Measurements check box and enter the Pipe Inner Diameter.<br />

See the Rosemount 5400 Reference Manual (Document No.<br />

00809-0100-4026) for more information on Hold Off Distance and other<br />

transmitter parameters.<br />

12. Click the Apply button to store the configuration.<br />

13. Select the Geometry tab.<br />

14. Enter the tank distance parameters.<br />

a. Tank Reference Height (R).<br />

The Tank Reference height (R) is defined as the distance from the<br />

Tank Reference Point to the Zero Level.<br />

b. Reference Distance (G).<br />

The Reference Distance (G) is the distance between the Tank<br />

Reference Point and the Gauge Reference Point, which is located at<br />

the top surface of the nozzle flange on which the gauge is mounted.<br />

5-82 Section 5. Installing a Raptor Level Gauging System