System Configuration Manual<br />

300510EN, Rev AA<br />

December 2010<br />

Rosemount Raptor<br />

9. To verify communication with the 2410 Tank Hub, type the current<br />

Modbus address and click the Verify Communication button. The Unit<br />

Id will appear if the correct Modbus address was entered (the 2410 is<br />

shipped with the default Modbus address=247).<br />

In case you would like to change the current Modbus address, or the<br />

address is unknown, click the Change Address on Device button.<br />

NOTE!<br />

In case several 2410 Tank Hubs are connected using the same default<br />

address (247), you will have to change the addresses before the Verify<br />

command can be used. See ”How to change the Modbus address of the<br />

2410” for more information.<br />

How to change the Modbus address of the 2410<br />

a. In the 2410 Tank Hub Communication window click the Change<br />

Address on Device... button to open the Change Address window.<br />

b. Enter the Unit ID and the new Modbus Address<br />

When changing the device address, the Unit Id is used as a unique<br />

identifier of the device. The Unit Id can be found on a label mounted<br />

on the device.<br />

Tip! If there is no other device connected that uses the same<br />

address as the current device, you can find the Unit Id by typing the<br />

current address into the Address field in the 2410 Tank Hub<br />

Communication window and clicking the Verify Communication<br />

button.<br />

c. Click the OK button to confirm the address settings and close the<br />

Change Address window.<br />

d. In the 2410 Tank Hub Communication window click the Verify<br />

Communication button to check that communication is established<br />

between the TankMaster work station and the 2410 Tank Hub. The<br />

Unit ID will appear when TankMaster finds the 2410.<br />

10. In the 2410 Tank Hub Communication window click the Next button to<br />

continue the installation procedure of the 2410.<br />

Section 5. Installing a Raptor Level Gauging System<br />

5-41