Sampling Guide for Air Contaminants in the Workplace - Irsst
Sampling Guide for Air Contaminants in the Workplace - Irsst
Sampling Guide for Air Contaminants in the Workplace - Irsst
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
ROHS Substances table<br />
119<br />
ROHS Name<br />
CAS<br />
TWAEV<br />
STEV<br />
Ceil<strong>in</strong>g<br />
(mg/m³)<br />
Notes<br />
<strong>Sampl<strong>in</strong>g</strong><br />
Device<br />
Flow<br />
L/m<strong>in</strong><br />
Volume<br />
TWAEV<br />
STEV<br />
(L)<br />
Pr<strong>in</strong>ciple<br />
MRV<br />
µg<br />
CVa<br />
(%)<br />
#<br />
meth.<br />
Remarks<br />
Stib<strong>in</strong>e (as Sb)<br />
7803-52-3<br />
0,51 Silica gel tube impregnated with mercuric 0,01- 50 UV/VIS-S<br />
NIOSH<br />
6008<br />
bichloride<br />
0,2<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
Stoddard solvent<br />
8052-41-3<br />
525 SKC 226-01 activated charcoal tube<br />
Max: 10 GC-FID 275 5,0 IRSST<br />
80-1<br />
2120 Activated charcoal 100/50 mg<br />
0,2<br />
A process sample must be supplied.<br />
Strontium chromate (as<br />
Cr)<br />
7789-06-2<br />
0,0005 C2<br />
RP<br />
EM<br />
PALL CA28149-306 v<strong>in</strong>yl/acrylic<br />
copolymer filter<br />
926 V<strong>in</strong>yl/acrylic copolymer filter<br />
1,5 à 4 360 IC-VIS 0,04 4,4 IRSST<br />
365<br />
For surface wipe sampl<strong>in</strong>g, contact Laboratory Customer Service.<br />
The analytical results are expressed as total water-<strong>in</strong>soluble chromium VI.<br />
Surface tests <strong>for</strong> chromates can be carried out, and <strong>the</strong> required equipment (kit<br />
#2625) is available at <strong>the</strong> IRSST. See communiqué 2008-06 on our Web site.<br />
Analysis time of 8 days <strong>for</strong> weld<strong>in</strong>g fumes and 10 days <strong>for</strong> <strong>the</strong> o<strong>the</strong>r processes<br />
follow<strong>in</strong>g <strong>the</strong> sampl<strong>in</strong>g date. Ship all samples by rapid courier with<strong>in</strong> a period not<br />
exceed<strong>in</strong>g 24 hours after sampl<strong>in</strong>g.<br />
Any substance that produces a response on <strong>the</strong> detector at 540 nm and that has<br />
a retention time similar to Cr VI is a potential <strong>in</strong>terference.<br />
Strychn<strong>in</strong>e<br />
57-24-9<br />
0,15 Glass fiber filter<br />
1-3 200 HPLC-UV<br />
NIOSH<br />
5016<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
Styrene (monomer)<br />
100-42-5<br />
213<br />
426<br />
Pc<br />
C3<br />
DRI-PAD<br />
IRSST<br />
39-A<br />
Refer to <strong>the</strong> user's guide accompany<strong>in</strong>g <strong>the</strong> <strong>in</strong>strument.<br />
Styrene is quantified <strong>in</strong> <strong>the</strong> field with a direct-read<strong>in</strong>g <strong>in</strong>strument (DRI). The<br />
m<strong>in</strong>imum reported value is of 0.5 mg/m³ (0.12 ppm). The reference method<br />
corresponds to calibration of <strong>the</strong> DRI.<br />
Styrene (monomer)<br />
100-42-5<br />
213<br />
426<br />
Pc<br />
C3<br />
3M Organic Vapor Monitor #3500<br />
2695 Passive dosimeter<br />
0,0289 GC-FID 27 1,4<br />
IRSST<br />
31-3<br />
For <strong>the</strong> monitors, <strong>the</strong> flow must be considered as be<strong>in</strong>g a sampl<strong>in</strong>g rate.<br />
Accord<strong>in</strong>g to <strong>the</strong> data from 3M, <strong>the</strong> precision (CVsampl<strong>in</strong>g) of <strong>the</strong> sampl<strong>in</strong>g rate is<br />
3.0%. Refer to <strong>the</strong> 3M passive monitor guide on <strong>the</strong> IRSST's Web site.<br />
Styrene (monomer)<br />
100-42-5<br />
213<br />
426<br />
Pc<br />
C3<br />
SKC 226-01 activated charcoal tube<br />
2120 Activated charcoal 100/50 mg<br />
Max:<br />
0,2<br />
5<br />
3<br />
GC-FID<br />
27 1,4<br />
IRSST<br />
31-3<br />
The analysis can be done by gas chromatography coupled with mass<br />
spectrometry (GC-MS, IRSST method #369); <strong>in</strong> this case <strong>the</strong> analytical<br />
uncerta<strong>in</strong>ty (CVa) is 3.0%.<br />
Subtilis<strong>in</strong>s (Proteolytic<br />
enzymes as 100 % pure<br />
crystall<strong>in</strong>e enzyme)<br />
--<br />
--<br />
0,00006<br />
RP<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
No support<strong>in</strong>g reference is suggested <strong>for</strong> this product.