Carbon dioxide<br />
124-38-9<br />
9000<br />
54000<br />
DRI-IR<br />
IRSST<br />
34-C<br />
Refer to <strong>the</strong> user's guide accompany<strong>in</strong>g <strong>the</strong> <strong>in</strong>strument.<br />
Carbon dioxide is quantified <strong>in</strong> <strong>the</strong> field with a direct-read<strong>in</strong>g <strong>in</strong>strument (DRI).<br />
The m<strong>in</strong>imum reported value is 18 mg/m³ (10 ppm). The reference method<br />
corresponds to calibration of <strong>the</strong> DRI.<br />
Carbon disulfide<br />
75-15-0<br />
12<br />
36<br />
Pc Activated charcoal tube preceded by a 0,01- 25<br />
dry<strong>in</strong>g tube<br />
0,2 3<br />
GC-FPD<br />
NIOSH<br />
1600<br />
The samples must be stored <strong>in</strong> <strong>the</strong> refrigerator until analysis. Contact <strong>the</strong> IRSST<br />
be<strong>for</strong>e sampl<strong>in</strong>g.<br />
Carbon monoxide<br />
630-08-0<br />
40<br />
230<br />
DRI-elec<br />
IRSST<br />
3-B<br />
Refer to <strong>the</strong> user's guide accompany<strong>in</strong>g <strong>the</strong> <strong>in</strong>strument.<br />
Quantification of carbon monoxide is done <strong>in</strong> <strong>the</strong> field with a direct-read<strong>in</strong>g<br />
<strong>in</strong>strument (DRI). The m<strong>in</strong>imum reported value is 1.1 mg/m³ (1 ppm). The<br />
reference method corresponds to calibration of <strong>the</strong> DRI.<br />
Carbon monoxide<br />
630-08-0<br />
40<br />
230<br />
DRI-PAD<br />
IRSST<br />
320-1<br />
The reported m<strong>in</strong>imum value is of 0.2 mg/m³ (0.2 ppm). In <strong>the</strong> case of a<br />
respirable compressed air (RCA) sample, <strong>the</strong> analysis is done at <strong>the</strong> IRSST.<br />
Carbon tetrabromide<br />
558-13-4<br />
1,4<br />
4,1<br />
XAD-4 tube<br />
0,2 10 GC-ECD<br />
3<br />
OSHA<br />
IMIS0565<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
Carbon tetrachloride<br />
56-23-5<br />
31<br />
63<br />
Pc<br />
C2<br />
EM<br />
3M 3500 organic vapour monitor<br />
2695 Passive dosimeter<br />
0,0302<br />
0<br />
GC-FID<br />
25 0,8<br />
IRSST<br />
157-2<br />
For <strong>the</strong> monitors, <strong>the</strong> flow must be considered as be<strong>in</strong>g a sampl<strong>in</strong>g rate.<br />
Accord<strong>in</strong>g to <strong>the</strong> data from 3M, <strong>the</strong> precision (CVsampl<strong>in</strong>g) of <strong>the</strong> sampl<strong>in</strong>g rate is<br />
1.3%. Refer to <strong>the</strong> 3M passive monitor guide on <strong>the</strong> IRSST's Web site.<br />
Carbon tetrachloride<br />
56-23-5<br />
31<br />
63<br />
Pc<br />
C2<br />
EM<br />
SKC 226-01 activated charcoal tube<br />
2120 Activated charcoal 100/50 mg<br />
Max:<br />
0,2<br />
15 GC-FID 25 0,8 IRSST<br />
157-2<br />
Carbonyl fluoride<br />
353-50-4<br />
5,4<br />
13<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
No support<strong>in</strong>g reference is suggested <strong>for</strong> this product.<br />
Catechol<br />
120-80-9<br />
23 Pc OVS-7 tube with XAD-7 and glass fiber 1 100 GC-FID<br />
OSHA<br />
PV2014<br />
filter<br />
Contact <strong>the</strong> IRSST be<strong>for</strong>e sampl<strong>in</strong>g.<br />
Cellulose (paper fibres)<br />
9004-34-6<br />
10 Pt Polyv<strong>in</strong>yl chloride filter with Accu-Cap. 1,5 180 Grav 25 4,9 IRSST<br />
48-1<br />
910 PVC filter-37 with Accu-Cap<br />
<strong>Sampl<strong>in</strong>g</strong> of respirable dusts is also possible by us<strong>in</strong>g <strong>the</strong> flow rate appropriate to<br />
<strong>the</strong> cyclone used.<br />
Additional <strong>in</strong><strong>for</strong>mation is available <strong>in</strong> Info-Labo 91-03. The weight measurement<br />
method <strong>for</strong> dusts is by def<strong>in</strong>ition non-specific.