Section 4 CPU and Power Supply<br />
4.6.2 Line Cord<br />
Replacement<br />
1. Feed the new line cord through the front of the mounting<br />
plate and tighten the black plastic nut onto the bulkhead<br />
fitting until both are flush against the plate.<br />
(The bulkhead fitting has a rubber ring that seals the box<br />
and prevents the fitting from coming loose.)<br />
2. Attach the ground wire to the mounting plate with the lock<br />
nut, and attach the blue and brown wires to their respective<br />
terminals (Figure 4-16).<br />
Note<br />
Before connecting, route these wires under those of the compressor<br />
cord, as shown in Figure 4-16, to prevent them from<br />
being caught under the edge of the protective cover.<br />
3. Ensure that the gasket in the cover is evenly seated in its<br />
groove. Align the cover over the eight mounting holes and<br />
reattach it with the self-tapping torx screws.<br />
CAUTION<br />
When reinstalling all self-tapping screws, avoid destroying the<br />
plastic threads. First seat each screw in its hole and, without<br />
pressing down, rotate the screw counter-clockwise until it falls<br />
into its thread groove with a "click." Then tighten the screw.<br />
4. Reinstall the power supply, following the instructions in<br />
Section 4.5.3.<br />
5. Close the rear compartment (see Section 1.8 Closing the<br />
Rear Compartment).<br />
4-25