Section 4 CPU and Power Supply<br />

4.6.2 Line Cord<br />

Replacement<br />

1. Feed the new line cord through the front of the mounting<br />

plate and tighten the black plastic nut onto the bulkhead<br />

fitting until both are flush against the plate.<br />

(The bulkhead fitting has a rubber ring that seals the box<br />

and prevents the fitting from coming loose.)<br />

2. Attach the ground wire to the mounting plate with the lock<br />

nut, and attach the blue and brown wires to their respective<br />

terminals (Figure 4-16).<br />

Note<br />

Before connecting, route these wires under those of the compressor<br />

cord, as shown in Figure 4-16, to prevent them from<br />

being caught under the edge of the protective cover.<br />

3. Ensure that the gasket in the cover is evenly seated in its<br />

groove. Align the cover over the eight mounting holes and<br />

reattach it with the self-tapping torx screws.<br />

CAUTION<br />

When reinstalling all self-tapping screws, avoid destroying the<br />

plastic threads. First seat each screw in its hole and, without<br />

pressing down, rotate the screw counter-clockwise until it falls<br />

into its thread groove with a "click." Then tighten the screw.<br />

4. Reinstall the power supply, following the instructions in<br />

Section 4.5.3.<br />

5. Close the rear compartment (see Section 1.8 Closing the<br />

Rear Compartment).<br />