5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />
Section 1 Introduction and General Maintenance<br />
1.2 Technical<br />
Specifications<br />
Table 1-1 contains technical specifications for the <strong>5800</strong> sampler.<br />
Table 1-1 <strong>5800</strong> Technical Specifications<br />
Size (HWD): 130 72 84 cm 52 29 33 in<br />
Weight (empty): 83.5 kg 184 lb<br />
Bottle Configurations:<br />
Refrigerator Body:<br />
Power Requirements:<br />
Installation Category:<br />
Pollution Degree: 3<br />
Maximum Altitude a :<br />
Humidity: 0 to 100%<br />
13 configurations available:<br />
24, 1-liter PP<br />
24, 350-ml glass<br />
4, 10-liter PE or glass<br />
4, 20-liter PE<br />
2, 10-liter PE or glass<br />
1, 20-liter PE or glass<br />
1, 10-liter PE or glass<br />
24 ProPaks, 1-liter wedge<br />
1 ProPak, 10-liter round<br />
Linear low-density polyethylene (LLDPE)<br />
115 VAC ±10%, 60 Hz <strong>Sampler</strong><br />
Running current 4 amperes<br />
Stopped current 22 amperes<br />
230 VAC ±10%, 50 Hz <strong>Sampler</strong><br />
Running current 2 amperes<br />
Stopped current 17 amperes<br />
II<br />
2,000 meters 6,562 feet<br />
Operational Temperature: –29 to 49 °C –20 to 120 °F<br />
Pump<br />
Intake Suction Tubing Length: 1 to 30 m 3 to 99 feet<br />
Material:<br />
Suction Line Inside Diameter:<br />
Pump Tubing Life:<br />
Vinyl or FEP-lined polyethylene<br />
9 mm<br />
Typically 1,000,000 pump counts.<br />
3 /8 inch<br />
Maximum Suction Lift: 8.5 m 28 feet<br />
Typical Repeatability<br />
(Ability to repeat the delivered volume<br />
for a set of samples collected under<br />
the same conditions.):<br />
Typical Delivered Volume Accuracy:<br />
±5ml or ±5% of the average volume in a set, whichever is greater, at lifts up to<br />
25 feet.<br />
±10ml or ±10% of programmed value, whichever is greater.<br />
Typical Line Velocity at Head Height: 0.9 m: 0.91 m/s 3 ft: 3.0 ft/s<br />
3.1 m: 0.87 m/s 10 ft: 2.9 ft/s<br />
4.6 m: 0.83 m/s 15 ft: 2.7 ft/s<br />
7.6 m: 0.67 m/s 25 ft: 2.2 ft/s<br />
Speed:<br />
Approx. 300 RPM, depending on sampling conditions<br />
1-2