Section 5 Electrical Troubleshooting and Diagnostics<br />
If the sampler displays PUMP FAILURE or BAD ON/OFF RATIO,<br />
turn off the sampler by pressing the Off button. Then, refer to<br />
Section 1.4.1 and perform a pump inspection. Clear any obstructions<br />
from the pump housing and pump tube, verify that the<br />
pump rotor turns, then perform the TEST PUMP diagnostic<br />
again. If you don’t find obstructions to pump rotor movement but<br />
the pump still fails, contact your authorized <strong>Isco</strong> service facility<br />
or the Teledyne <strong>Isco</strong> factory.<br />
5.1.6 Test Distributor The TEST DISTRIBUTOR diagnostic calibrates the sampler’s distributor<br />
alignment. You specify the bottle position to which the<br />
sampler should move the arm. You can then verify the correct<br />
positioning of the arm.<br />
Note<br />
This test is not applicable for samplers configured for operation<br />
with a single 20 liter bottle.<br />
To run the test:<br />
1. Remove the discharge tube from the distributor arm.<br />
This removes tension from the arm for better visual verification.<br />
C<br />
A<br />
B<br />
Figure 5-1 Distributor arm (A), discharge tube (B), and stop (C)<br />
SELECT DIAG: ()<br />
TEST DISTRIBUTOR<br />
2. Go to the RUN DIAGNOSTICS menu as usual and select the<br />
TEST DISTRIBUTOR option.<br />
3. Open the refrigerated compartment and manually move<br />
the distributor arm to its “home” position against the front<br />
of the stop on the distributor module (Figure 5-2).<br />
5-4