Section 5 Electrical Troubleshooting and Diagnostics<br />
· poor contrast<br />
If any of these or other faulty display conditions exist, contact<br />
your authorized <strong>Isco</strong> service facility or the Teledyne <strong>Isco</strong> factory.<br />
5.1.4 Test Keypad The TEST KEYPAD diagnostic tests the sampler’s keypad. While<br />
the test is active, the display reports the button name of any<br />
button you press. There is no pass or fail message. The technician<br />
must watch the display to determine if the sampler<br />
reports the correct button when it is pressed.<br />
To run the test:<br />
SELECT DIAG: ()<br />
TEST KEYPAD<br />
PRESS ANY KEY<br />
WATCH DISPLAY<br />
ENTER<br />
1. Select the TEST KEYPAD option and press Enter.<br />
2. The sampler displays the test instructions.<br />
3. Press any buttons on the keypad in any order; the display<br />
should report the button name as it is pressed. For example,<br />
press the Enter button. The display reports ENTER.<br />
After 20 keystrokes, the sampler reverts to the SELECT<br />
DIAG menu.<br />
Note<br />
Pressing the red Stop button twice consecutively will exit the<br />
test regardless of the number of strokes.<br />
If the display does not report the button name when you press it,<br />
the keypad could be faulty. Contact your authorized <strong>Isco</strong> service<br />
facility or the Teledyne <strong>Isco</strong> factory.<br />
5.1.5 Test Pump The TEST PUMP diagnostic tests the sampler’s pump assembly.<br />
The sampler operates the pump in each direction and reports an<br />
ON/OFF ratio. This ratio is an indication of the pump’s<br />
mechanical and electronic performance.<br />
To run the test:<br />
SELECT DIAG: ()<br />
TEST PUMP<br />
1. Select the TEST PUMP option and press Enter. The sampler<br />
starts the test and continues without intervention. To<br />
cancel the test you can press the Stop button.<br />
PUMPING...<br />
PUMPING...<br />
ON/OFF RATIO = ___<br />
2. The sampler displays the On/Off ratio. To pass, the number<br />
must be between 0.8 and 1.25. The test continues by operating<br />
the pump in reverse (purge).<br />
PURGING...<br />
PURGING...<br />
ON/OFF RATIO = ___<br />
3. The sampler displays the On/Off ratio. To pass, the number<br />
must be between 0.8 and 1.25.<br />