07.03.2014 Views

5800 Refrigerated Sampler Service Guide - Isco

5800 Refrigerated Sampler Service Guide - Isco

5800 Refrigerated Sampler Service Guide - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 3 Refrigeration System<br />

a. For 230V units, loosen the screw on the top hold-down<br />

tab to free the harness.<br />

Smooth<br />

Ribbed<br />

115V<br />

230V<br />

Top hold-down<br />

tab<br />

Ribbed<br />

Smooth<br />

Figure 3-14 Condenser fan connections<br />

3. Remove the three lock nuts holding the fan assembly in<br />

place. Remove the fan assembly and vibration cushion.<br />

Lock<br />

nuts (3)<br />

Fan<br />

Drip<br />

tray<br />

Vibration<br />

cushion<br />

Figure 3-15 Removal/replacement of condenser fan<br />

4. Place the new vibration cushion and fan assembly on the<br />

floor of the refrigeration module, aligning the holes in the<br />

base over the threaded mounting studs, and secure with<br />

the three lock nuts.<br />

5. Route the fan harness around the drip tray and connect to<br />

the relay, taking care to connect the ribbed and smooth<br />

wires correctly, as shown in Figure 3-14.<br />

a. For 230V units, also route the harness wires around<br />

the right side of the terminal box and under the top<br />

hold-down tab. Tighten the screw.<br />

6. Replace the terminal cover, taking care not to pinch or bind<br />

any of the wires.<br />

3-18

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!