Section 5 Electrical Troubleshooting and Diagnostics<br />

5.1.7 Test Flow Meter Port The TEST FLOWMETER PORT diagnostic tests the sampler’s<br />

ability to send event marks, and receive an enable pin signal and<br />

flow pacing pulses.<br />

Note<br />

This diagnostic routine is not for field use. This test requires a<br />

test plug which is inserted into the External Device connector.<br />

If you have checked all the connections with the external flow<br />

meter and still suspect a fault with the event mark or flow pacing<br />

signals, contact your authorized <strong>Isco</strong> service facility or the<br />

Teledyne <strong>Isco</strong> factory. They will be able to assist you with<br />

obtaining a test plug or making the necessary jumper connections<br />

to run the test.<br />

SELECT DIAG: ()<br />

TEST FLOWMETER PORT<br />

INSTALL TEST PLUG<br />

PRESS WHEN READY!<br />

FLOWMETER PORT TEST:<br />

...TESTING<br />

To run the test:<br />

1. Select the TEST FLOWMETER PORT option and press<br />

Enter.<br />

2. Insert a Flow Meter Port Test Plug into the Flow Meter<br />

connector.<br />

3. The sampler tests the send and receive circuits for the<br />

event marks, sampler enable pin (inhibit), and flow pulses.<br />

If the sampler displays FAILED: EM->FP or FAILED: INHIBIT,<br />

verify that test plug is correctly inserted. Then, perform the<br />

TEST FLOWMETER PORT diagnostic again from step 1. If the<br />

test still fails, contact your authorized <strong>Isco</strong> service facility or the<br />

Teledyne <strong>Isco</strong> factory.<br />

5.1.8 Refrig Temperature The REFRIG TEMPERATURE diagnostic displays the temperature<br />

of the refrigerated compartment. When this test is started,<br />

the sampler should display the temperature until you press the<br />

Stop or Enter button. There is no pass or fail. This test simply<br />

provides continuous temperature monitoring.<br />

If the screen displays an asterisk (*) or inaccurate temperature<br />

reading, the temperature sensor cable may be malfunctioning.<br />

See Section 3.3.1 for replacement procedures.<br />

SELECT DIAG: ()<br />

REFRIG TEMPERATURE<br />

REFRIG TEMPERATURE:<br />

AIR=__C EVAP=__C<br />

To display the refrigerator temperature:<br />

1. Select the REFRIG TEMPERATURE option and press Enter.<br />

2. The display shows the temperature until you press the<br />

Stop or Enter button.<br />

5-6