Section 2 Sample Delivery System<br />
G<br />
W<br />
B<br />
R<br />
Yel.<br />
B/R<br />
R/B tw.<br />
Figure 2-3 Pump board connections<br />
2. Align the module over the mounting holes and secure it<br />
with the mounting screws, inserting the top three first.<br />
CAUTION<br />
Ensure that all wires are tucked into the pump module compartment<br />
and not caught between the gasket and mounting<br />
surface.<br />
Insert top row<br />
of screws first.<br />
Ensure that wires<br />
are not caught<br />
under gasket.<br />
Figure 2-4 Installing the pump module<br />
3. After all six screws are inserted, torque the screws 1.8 to<br />
2.0 Nm (16 to 18 in-lbs).<br />
4. Install the pump tube.<br />
a. Loosen the Liquid Detector Cover (F) by unscrewing<br />
the large black knob (H).<br />
b. Unlatch the Pump Housing Band (J).<br />
c. Thread the pump tube through the pump. Note that<br />
the short end (inlet) should extend from the top opening.<br />
2-4