Section 2 Sample Delivery System<br />

2.3.8 Gasket Replace the gasket using the following steps, as shown in<br />

Figure 2-17.<br />

1. Always begin by pressing the gasket into one corner of the<br />

base.<br />

2. Pull the gasket taut, then press the gasket down into the<br />

opposite corner. This prevents the gasket from being<br />

unevenly distributed in the groove, ensuring the integrity<br />

of the seal.<br />

3. Press the other corners in, followed by the rest of the gasket.<br />

Figure 2-17 Proper installation of pump gasket<br />

Note<br />

Installing the gasket without following these steps will result in<br />

uneven distribution of the gasket, as shown in Figure 2-18.<br />

This will result in moisture/liquids entering the motor housing,<br />

leading in turn to internal damage to the pump module.<br />

2-13