Section 2 Sample Delivery System<br />
2.3.8 Gasket Replace the gasket using the following steps, as shown in<br />
Figure 2-17.<br />
1. Always begin by pressing the gasket into one corner of the<br />
base.<br />
2. Pull the gasket taut, then press the gasket down into the<br />
opposite corner. This prevents the gasket from being<br />
unevenly distributed in the groove, ensuring the integrity<br />
of the seal.<br />
3. Press the other corners in, followed by the rest of the gasket.<br />
Figure 2-17 Proper installation of pump gasket<br />
Note<br />
Installing the gasket without following these steps will result in<br />
uneven distribution of the gasket, as shown in Figure 2-18.<br />
This will result in moisture/liquids entering the motor housing,<br />
leading in turn to internal damage to the pump module.<br />
2-13