<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 2 Sample Delivery System<br />

1.<br />

3.<br />

2.<br />

Figure 2-28 Distributor gasket replacement<br />

Note<br />

Installing the gasket without following these steps will result in<br />

uneven distribution of the gasket, as shown in Figure 2-18.<br />

This will result in moisture/liquids entering the motor housing,<br />

leading in turn to internal damage to the pump module.<br />

2.8.2 Distributor Shaft To assemble the distributor shaft, refer to Figure 2-29 and do the<br />

following:<br />

1. Place the disk spacer on the bottom side (shaft side) of the<br />

shaft gear, aligning the four pairs of holes.<br />

2. Position the index disk so that the notch in the disk is visible<br />

and aligned with the notch in the distributor shaft.<br />

Replace the four screws.<br />

3. Reinstall the distributor shaft in the brass bearing in the<br />

base.<br />