5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />
Section 2 Sample Delivery System<br />
1.<br />
3.<br />
2.<br />
Figure 2-28 Distributor gasket replacement<br />
Note<br />
Installing the gasket without following these steps will result in<br />
uneven distribution of the gasket, as shown in Figure 2-18.<br />
This will result in moisture/liquids entering the motor housing,<br />
leading in turn to internal damage to the pump module.<br />
2.8.2 Distributor Shaft To assemble the distributor shaft, refer to Figure 2-29 and do the<br />
following:<br />
1. Place the disk spacer on the bottom side (shaft side) of the<br />
shaft gear, aligning the four pairs of holes.<br />
2. Position the index disk so that the notch in the disk is visible<br />
and aligned with the notch in the distributor shaft.<br />
Replace the four screws.<br />
3. Reinstall the distributor shaft in the brass bearing in the<br />
base.<br />
2-23