Section 4 CPU and Power Supply<br />

Figure 4-4 Remove adapter from power cord<br />

b. Slide the bushing sideways away from the refrigeration<br />

module until it is free (Figure 4-5). Use care to avoid<br />

bending the refrigeration tubing.<br />

Figure 4-5 Remove power cord<br />

c. Pull the power cord through the refrigeration module.<br />

2. Remove the electrical cover on the side of the compressor.<br />

Disconnect the blue and brown wires from the spade connections<br />

on the side of the compressor, and the green and<br />

yellow ground wire from the module chassis. (These three<br />

wires are from the Power Supply module.)<br />

· For 115 VAC systems, refer to Figure 4-6.<br />

· For 230 VAC systems, refer to Figure 4-7.<br />

3. Open the top cover of the sampler and remove the control<br />

panel mounting screws (see Figure 4-8).<br />

4. Lift the control panel away from the refrigerator body to<br />

expose the wiring connectors.<br />

5. Disconnect wiring connectors P1 and P10.<br />

6. Place the control panel back on the refrigerator body and<br />

start at least two screws to temporarily hold the panel<br />

until the new harness is in place.<br />

4-16