Section 2 Sample Delivery System<br />
2. Place the stepper motor on the chassis without tightening<br />
the mounting screws, ensuring that the belt is around the<br />
motor gear.<br />
3. Place the cover plate on top of the shaft and reattach with<br />
the three large self-tapping screws.<br />
Figure 2-31 Installing the cover plate<br />
4. While maintaining the belt tension by pressing the motor<br />
away from the distributor shaft, slowly turn the large gear<br />
until you feel the belt jerk as the teeth fall into alignment<br />
(Figure 2-32).<br />
Turn the large gear at least one full revolution to ensure<br />
that all teeth have engaged between the pulley and the<br />
belt. This eliminates the gaps shown in Figure 2-30.<br />
5. Still pressing back on the motor, tighten the two motor<br />
plate mounting screws.<br />
Press &<br />
turn<br />
Press<br />
back<br />
Press &<br />
turn<br />
Figure 2-32 Timing belt alignment<br />
2-25