Section 2 Sample Delivery System<br />

2. Place the stepper motor on the chassis without tightening<br />

the mounting screws, ensuring that the belt is around the<br />

motor gear.<br />

3. Place the cover plate on top of the shaft and reattach with<br />

the three large self-tapping screws.<br />

Figure 2-31 Installing the cover plate<br />

4. While maintaining the belt tension by pressing the motor<br />

away from the distributor shaft, slowly turn the large gear<br />

until you feel the belt jerk as the teeth fall into alignment<br />

(Figure 2-32).<br />

Turn the large gear at least one full revolution to ensure<br />

that all teeth have engaged between the pulley and the<br />

belt. This eliminates the gaps shown in Figure 2-30.<br />

5. Still pressing back on the motor, tighten the two motor<br />

plate mounting screws.<br />