Section 2 Sample Delivery System<br />

Figure 2-18 Improper installation of pump gasket<br />

2.3.9 Detector Protective<br />

Barrier<br />

The liquid detector’s protective fluoroglass barrier (Figure 2-19)<br />

can be removed for cleaning or replacement. It is anchored by a<br />

hold-down panel on either side.<br />

Note<br />

The piezo film liquid detector is permanently attached to the<br />

pump base. If it is damaged, the entire pump base must be<br />

replaced.<br />

Barrier removal and<br />

replacement<br />

1. Remove the four self-tapping Phillips screws attaching<br />

the hold-down panels and pull them off.<br />

2. Peel the barrier away from the detector.<br />

3. Apply a thin layer of 737 RTV sealant (800-1002-03) to the<br />

outer edges of the mounting surface, taking care to avoid<br />

getting any in the tubing grooves.<br />

4. Install the new barrier, fitting the four inside holes over<br />

the insert guides, as shown in Figure 2-19.<br />

5. Reattach the two hold-down panels, pressing them down<br />

over the insert guides. Torque the self-tapping screws to<br />

7 - 10 in/lbs.<br />

2-14