<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 3 Refrigeration System<br />

3.4.5 Module Replacement The refrigeration module is designed to be replaced as a complete<br />

unit.<br />

Unlike the other modules, the refrigeration module is<br />

voltage-specific. Be sure to order the replacement module and/or<br />

appropriate components for your AC mains power. Table 3-1 on<br />

page 3-1 contains part numbers for the refrigeration system and<br />

its components. For a complete listing of replacement parts, refer<br />

to Appendix B.<br />

Required Parts and Tools • Replacement refrigeration module<br />

• Nylon cable tie 489-0110-00 (For best results, soak in<br />

water for 1 hour prior to use.)<br />

• #2 & #3 Phillips screwdrivers<br />

• 15 /16" open wrench<br />

• 11 /32" open wrench<br />

• Wire cutters<br />

CAUTION<br />

Replacement of individual components may compromise the<br />

closed refrigerant system and the protective coatings.<br />

Removal<br />

1. Cut the cable tie holding the power cord that runs through<br />

the refrigeration assembly.<br />

Cut cable<br />

tie<br />

Figure 3-16 Cut cable tie to free power cord (Full power<br />

cord length not depicted)<br />

2. Remove the power cord from the refrigeration assembly.<br />

a. Slide the bushing sideways away from the refrigeration<br />

module until it is free. Use care to avoid bending the<br />

refrigeration tubing.<br />

b. Pull the power cord through the module.<br />

c. 230V/50 Hz Systems Only — Remove the AC Plug<br />

adapter from the end of the power cord by first cutting<br />

away the plastic heat shrink tubing. Then pull the<br />

adapter off, leaving just the North American 115V plug.<br />

3-19