Section 1 Introduction and General Maintenance<br />
4. Inspect the following:<br />
a. Pump tube – Look for excessive wear caused by the<br />
pump rollers and for cracks along the sides of the tube.<br />
Cracks might not be apparent unless you flex or<br />
squeeze the tubing. If cracks or excessive wear are evident,<br />
replace the pump tube (Section 1.4.2).<br />
G<br />
D<br />
A<br />
E<br />
F<br />
H<br />
I<br />
J<br />
C<br />
B<br />
Figure 1-2<br />
<strong>5800</strong> Pump<br />
A. Pump Tube F. Liquid Detector Cover<br />
B. Bulkhead Fitting G. Latch<br />
C. Tubing Coupler H. Knob<br />
D. Alignment Notches I. Pump Housing<br />
E. Alignment Collars J. Pump Housing Band<br />
CAUTION<br />
This sampler has a high performance pump. As with all such<br />
pumps, it relies upon liquid to cool working components. If the<br />
sampler is programmed to pump in the absence of liquid in<br />
excess of 5 minutes, excessive heat buildup may damage the<br />
paddles, rollers, and housing. Ensure that the liquid inlet is<br />
completely immersed.<br />
1-6