<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 1 Introduction and General Maintenance<br />

4. Inspect the following:<br />

a. Pump tube – Look for excessive wear caused by the<br />

pump rollers and for cracks along the sides of the tube.<br />

Cracks might not be apparent unless you flex or<br />

squeeze the tubing. If cracks or excessive wear are evident,<br />

replace the pump tube (Section 1.4.2).<br />

G<br />

D<br />

A<br />

E<br />

F<br />

H<br />

I<br />

J<br />

C<br />

B<br />

Figure 1-2<br />

<strong>5800</strong> Pump<br />

A. Pump Tube F. Liquid Detector Cover<br />

B. Bulkhead Fitting G. Latch<br />

C. Tubing Coupler H. Knob<br />

D. Alignment Notches I. Pump Housing<br />

E. Alignment Collars J. Pump Housing Band<br />

CAUTION<br />

This sampler has a high performance pump. As with all such<br />

pumps, it relies upon liquid to cool working components. If the<br />

sampler is programmed to pump in the absence of liquid in<br />

excess of 5 minutes, excessive heat buildup may damage the<br />

paddles, rollers, and housing. Ensure that the liquid inlet is<br />

completely immersed.<br />

1-6