Section 2 Sample Delivery System<br />
2.1 Pump To produce accurate and repeatable samples, the pump uses a<br />
pump revolution counter and patented 1 non-contacting liquid<br />
detector. The <strong>5800</strong> uses a peristaltic pump housed in a separate,<br />
easily serviceable and fully replaceable module.<br />
CAUTION<br />
This sampler has a high performance pump. As with all such<br />
pumps, it relies upon liquid to cool working components. If the<br />
sampler is programmed to pump in the absence of liquid in<br />
excess of 5 minutes, excessive heat buildup may damage the<br />
paddles, rollers, and housing. Ensure that the liquid inlet is<br />
completely immersed.<br />
Liquid Present .75V<br />
Functional testing and<br />
diags<br />
Test points<br />
For functional CPU board testing through Hyper Terminal, see<br />
Section 4.3.3 Functional Tests. The tests are listed by category in<br />
Table 4-2.<br />
Refer to Section 5.1.5 for standard <strong>5800</strong> pump diagnostics.<br />
For a complete list of <strong>5800</strong> test points and values, refer to<br />
Section 4.2. Test points are also highlighted in red on the CPU<br />
schematic in Figure A-4, Appendix A.<br />
Test Point 5, located on pin 5 of U4-A on the CPU, gives the filtered<br />
AC signal from the liquid detector.<br />
Test Point 1 is the liquid detector rectified A to D signal, or the<br />
input to CR2. When liquid is present, this value should be<br />
greater than 0.75V, as shown at left.<br />
No Liquid Present 0V<br />
1. United States patent 5,125,801<br />
2-1