Section 1 Introduction and General Maintenance<br />
5. Start Flowlink’s application Update Software.<br />
a. From the computer START menu, select PROGRAMS,<br />
and highlight Flowlink.<br />
b. Select Update Software.<br />
6. Follow the update instructions in the Update Software<br />
Help menu.<br />
1.5.2 Software Reset If the <strong>5800</strong> does not appear to be operating normally, you can<br />
restart the processor by turning the unit off, waiting a few<br />
minutes, and then turning the unit back on. If the problem<br />
recurs, you can reset the software.<br />
CAUTION<br />
This procedure will cause most programmed entries and accumulated<br />
data stored in the sampler to be lost, and the sampler<br />
will revert to factory default settings. If this operation is performed,<br />
it will be necessary for you to reprogram the unit to<br />
meet the specifications of your installation. Record your program<br />
settings and download any data before performing a software<br />
reset.<br />
To reset the software:<br />
1. Disconnect power to the sampler.<br />
2. Hold down the Stop and Enter buttons together and reconnect<br />
the power.<br />
1.5.3 System IDs Select this option to view the system IDs. This function reports<br />
the unique ID for the sampler, and its hardware and software<br />
versions. These IDs are factory set.<br />
SELECT OPTION: ()<br />
SYSTEM IDs<br />
<strong>5800</strong> <strong>Sampler</strong><br />
ID: ___________<br />
HARDWARE: ___<br />
SOFTWARE: _._<br />
1. Select the SYSTEM IDs option and press Enter.<br />
2. The first line lists the model number. The second line lists<br />
the unique ID for the sampler. Press Enter to continue.<br />
This screen lists the version numbers of the installed hardware<br />
and software. Press Enter to return to the<br />
SELECT OPTION screen.<br />
1-12