<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 1 Introduction and General Maintenance<br />

1.6 Major Assembly<br />

Removal and<br />

Replacement<br />

Teledyne <strong>Isco</strong> designed the sampler for easy field replacement of<br />

its sub-assemblies. This expedites the repair and saves shipping<br />

costs. These sub-assembly modules should be replaced by trained<br />

personnel.<br />

The sub-assemblies are:<br />

• Pump Module (Section 2.2)<br />

• Distributor Module (Section 2.6)<br />

• Refrigeration Module (Section 3.4.5)<br />

• Control Panel Module with Main CPU board<br />

(Section 4.4)<br />

• Power Supply (Section 4.5)<br />

The removal and replacement instructions include general steps<br />

that apply to all modules, and detailed instructions for each<br />

module.<br />

Before proceeding with the detailed instructions in the following<br />

sections, familiarize yourself with these general steps for the<br />

removal and replacement of a module.<br />

1.6.1 Removing a Module 1. Unplug the AC power cord.<br />

2. Remove the mounting screws.<br />

3. Carefully lift the module away from the refrigerator body.<br />

4. Label the wires with the location of their connection, then<br />

disconnect the wiring connectors.<br />

1.6.2 Replacing a Module 1. Ensure that the gasket seal is in place. The gasket surface<br />

must be smooth, clean, and free of nicks, cracks, etc. If any<br />

imperfections are visible, replace the seal.<br />

2. Reconnect the wiring connectors, using care to correctly<br />

pair the connectors and to correctly align the keys and<br />

pins.<br />

3. Ensuring that no wires are caught between the module<br />

and mounting surface, align the module over the mounting<br />

holes and secure it with the mounting screws.<br />

4. Apply AC power and turn on the sampler.<br />

5. After replacement, test the new module using the appropriate<br />

diagnostic tests in Section 5.1.<br />

1.6.3 Safety Precautions There are several precautions you should observe before<br />

attempting to remove and replace a module:<br />

WARNING<br />

Removing a module exposes you to electrical and<br />

mechanical hazards. Always unplug the AC power cord<br />

before attempting to remove any module. Only trained<br />

service personnel may remove or replace these modules.<br />

1-13