<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />
Section 1 Introduction and General Maintenance<br />
1.6 Major Assembly<br />
Removal and<br />
Replacement<br />
Teledyne <strong>Isco</strong> designed the sampler for easy field replacement of<br />
its sub-assemblies. This expedites the repair and saves shipping<br />
costs. These sub-assembly modules should be replaced by trained<br />
personnel.<br />
The sub-assemblies are:<br />
• Pump Module (Section 2.2)<br />
• Distributor Module (Section 2.6)<br />
• Refrigeration Module (Section 3.4.5)<br />
• Control Panel Module with Main CPU board<br />
(Section 4.4)<br />
• Power Supply (Section 4.5)<br />
The removal and replacement instructions include general steps<br />
that apply to all modules, and detailed instructions for each<br />
module.<br />
Before proceeding with the detailed instructions in the following<br />
sections, familiarize yourself with these general steps for the<br />
removal and replacement of a module.<br />
1.6.1 Removing a Module 1. Unplug the AC power cord.<br />
2. Remove the mounting screws.<br />
3. Carefully lift the module away from the refrigerator body.<br />
4. Label the wires with the location of their connection, then<br />
disconnect the wiring connectors.<br />
1.6.2 Replacing a Module 1. Ensure that the gasket seal is in place. The gasket surface<br />
must be smooth, clean, and free of nicks, cracks, etc. If any<br />
imperfections are visible, replace the seal.<br />
2. Reconnect the wiring connectors, using care to correctly<br />
pair the connectors and to correctly align the keys and<br />
pins.<br />
3. Ensuring that no wires are caught between the module<br />
and mounting surface, align the module over the mounting<br />
holes and secure it with the mounting screws.<br />
4. Apply AC power and turn on the sampler.<br />
5. After replacement, test the new module using the appropriate<br />
diagnostic tests in Section 5.1.<br />
1.6.3 Safety Precautions There are several precautions you should observe before<br />
attempting to remove and replace a module:<br />
WARNING<br />
Removing a module exposes you to electrical and<br />
mechanical hazards. Always unplug the AC power cord<br />
before attempting to remove any module. Only trained<br />
service personnel may remove or replace these modules.<br />
1-13