<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 2 Sample Delivery System<br />

Latch screws<br />

Figure 2-10 Pump band latch<br />

2.3.4 Optical Disk & Sensor Removal of the optical disk may become necessary, either for<br />

replacement of the disk and sensor, or in order to free the drive<br />

shaft.<br />

Tools<br />

#2 Phillips screwdriver, 11 /32" nut driver, 3 /32" Allen wrench<br />

To remove the optical disk, grasp the pump rotor to hold the<br />

shaft in place and remove the mounting nut. Carefully tilt and<br />

slide the disk away from the sensor. If the optical sensor must be<br />

removed, remove the two Allen screws holding it in place.<br />

Remove<br />

nut<br />

Remove<br />

2 Allen<br />

screws<br />

Figure 2-11 Optical disk and sensor removal<br />

CAUTION<br />

Do not allow the disk to scrape against the optical sensor during<br />

removal/replacement.<br />

The optical sensor is fragile. If it does not need to be replaced,<br />

set it aside so that it does not get broken.<br />