5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />
Section 2 Sample Delivery System<br />
Figure 2-21 Install heat shield inside cover<br />
3. Temporarily close the pump cover to check the heat shield<br />
alignment. The hole in the shield must fit over the large<br />
knob on the pump assembly. If the shield hits the knob,<br />
loosen the four screws and adjust the shield. When the<br />
alignment is correct, tighten the screws.<br />
4. Position the heater on the sampler as shown in<br />
Figure 2-22. Align the edges of the slot in the heater<br />
against the lower left corner of the pump assembly.<br />
Figure 2-22 Install heater under pump<br />
5. Secure the heater in place with the two #142" self-tapping<br />
screws. Tighten the screws until fully seated.<br />
6. Refer to Figure 2-23 and secure the AC line cord just inside<br />
the notch in the cover, with the clip and the #8 5 /8"<br />
self-tapping screw.<br />
2-17