07.03.2014 Views

5800 Refrigerated Sampler Service Guide - Isco

5800 Refrigerated Sampler Service Guide - Isco

5800 Refrigerated Sampler Service Guide - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />

Section 3 Refrigeration System<br />

Figure 3-17 Removing 230V adaptor from AC power cord<br />

Bushing<br />

Figure 3-18 Remove power cord<br />

3. Gently remove the terminal cover of the compressor. Prying<br />

the top edge off with a screwdriver will release the<br />

cover.<br />

4. Disconnect the solid brown wire from the compressor terminals.<br />

Disconnect the blue wire from the other blue wire.<br />

5. With the 11 /32" wrench, remove the nut holding the green<br />

and yellow ground wire on the module chassis<br />

• For 115V systems, refer to Figure 3-19.<br />

• For 230V systems, refer to Figure 3-20.<br />

115V<br />

Brown<br />

Blue<br />

Green/<br />

Yellow<br />

Figure 3-19 Electrical connections on 115V compressor<br />