5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
5800 Refrigerated Sampler Service Guide - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>5800</strong> <strong>Refrigerated</strong> <strong>Sampler</strong> <strong>Service</strong> <strong>Guide</strong><br />
Section 3 Refrigeration System<br />
Figure 3-17 Removing 230V adaptor from AC power cord<br />
Bushing<br />
Figure 3-18 Remove power cord<br />
3. Gently remove the terminal cover of the compressor. Prying<br />
the top edge off with a screwdriver will release the<br />
cover.<br />
4. Disconnect the solid brown wire from the compressor terminals.<br />
Disconnect the blue wire from the other blue wire.<br />
5. With the 11 /32" wrench, remove the nut holding the green<br />
and yellow ground wire on the module chassis<br />
• For 115V systems, refer to Figure 3-19.<br />
• For 230V systems, refer to Figure 3-20.<br />
115V<br />
Brown<br />
Blue<br />
Green/<br />
Yellow<br />
Figure 3-19 Electrical connections on 115V compressor<br />
3-20