6 Checking and Adjustment<br />

Checking and Adjusting the Plate Level 6.1<br />

The axis of the plate level vial must be at right angles to the vertical axis of the<br />

instrument.<br />

To check and adjust the plate level:<br />

1. Set up the instrument on the tripod.<br />

2. Follow the leveling procedures described in Leveling, page 16.<br />

3. Rotate the alidade 180°.<br />

4. Check whether the bubble is in the center of the vial.<br />

5. If the bubble is not in the center of the vial, adjust the plate level:<br />

a. Using the adjusting pin supplied, rotate<br />

the plate level’s adjustment screw until<br />

the bubble has moved half of the<br />

distance back to the center.<br />

b. Using leveling screw A, move the<br />

bubble into the center of the vial.<br />

c. Repeat from Step 4.<br />

A<br />

B<br />

C<br />

2<br />

Checking and Adjusting the Circular Level 6.1<br />

Once you have checked and adjusted the plate<br />

level, check the circular level.<br />

If the bubble is not in the center of the level,<br />

use the adjusting pin to rotate the three<br />

adjustment screws until the bubble is centered.<br />

170 Pulse Laser Station <strong>NPL</strong>-302 Series Instruction <strong>Manual</strong>