Fate of Mercury in the Arctic 46<br />

concentration in a flux system sampling in three channels: up drafts, down drafts and all else, as<br />

compared to levels measured in parallel annular denuders under constant sampling conditions,<br />

could be set at 1 Hz.<br />

In these experiments flow rates were controlled by a Tylan mass flow controller and checked<br />

prior to and after the sampling period. The results of these experiments led to the sampling<br />

switching rate being set in the REA system to 1 Hz with the 10 litre per minute flow rate maintained<br />

by a mass flow controller. The air sampling-switching rate is so long compared to errors that could<br />

be induced from physical switching and software, that these are considered negligible. Zero air was<br />

not added to the sampling inlet system since this would mean that ambient air could not be directly<br />

sampled into the annular denuder, but that ambient air would have to flow through an additional<br />

inlet first. RGM is so reactive that it was thought best to directly sample it into the annular<br />

denuders. This is possible given the 1 Hz switching rate.<br />

Therefore, by sampling at 1 Hz, approximately 95% of the turbulence is captured and the best<br />

compromise between the meteorological measurements and chemical sampling is obtained in order<br />

to ensure a laminar flow in the annular denuders and thereby measure RGM flux most accurately<br />

Measurements were made as follows: The heating mantles are turned on approximately 1 hour<br />

prior to measurement. 4 annular denuders were prepared and blanked immediately prior to use: 3<br />

for the system and one as the field blank. Inlets are baked prior to field deployment unless<br />

contaminated by handling. Powder-free rubber gloves are used when handling the inlets. The<br />

denuders are then taken outdoors and carefully placed into the heating caps. Care is taken not to<br />

disturb the sonic alignment, or crack a denuder due to handling, nor handle the sampling inlet since<br />

the annular denuders are relatively fragile under Arctic weather conditions, and the inlets may<br />

easily be contaminated. Flow is checked in the sampling lines and temperature measured in the<br />

denuders. Once the temperature of the denuders has stabilized, the REA system is started by