The “Static Automatic” compensation:Consult the Help file, keyword index:Static Manual (Ohmic drop compensation).The “Dynamic” compensation:Consult the Help file, keyword index:Dynamic (Ohmic drop compensation).4.10.3 Table of the R min/R max limits that can be compensateddepending on the type of compensation performed and thecurrent range in useCurrent range R min (in Ohms) R min (in Ohms) R min (in Ohms)R max (in Ohms) R max (in Ohms) R max (in Ohms)Feed back Auto & Manual Static Auto & Manual Dynamic1 µA 244 - 1 000 000 0 - 10 000 000 33300 - 4 000 00010 µA 24.4 - 100 000 0 - 1 000 000 3330 - 400 000100 µA 2.44 - 10 000 0 - 100 000 333 - 40 0001 mA 0.244 - 1000 0 - 10 000 33.3 - 400010 mA 0.0244 - 100 0 - 1000 3.33 - 400100 mA 0 - 10 0 - 100 0.333 - 401 A 0 - 1 0 - 10 Not possibleAutoranging Not possible 0 - 10 000 000 0.333 - 2 000 000Max scan rate 10 V/s 1 V/s 100 mV/sVoltaMaster 4 - Version 7.0x 101Section 4, Chapter 10: The ohmic drop compensation