1 PGZ/PST050 “REF.” socket to the reference electrode A95A122(BNC/1.00 m/FX-M)2 PGZ/PST050 “WORK” socket to the EDI101 banana socketCable (UHF-PL259/1.00 m/FX-M)EDI Adapter 5B4mm-F/B2mm-M)A95A123X41V00113 PGZ/PST050 “AUX.” socket to the auxiliary electrode A95A121(UHF-PL259/1.00 m/FX-M)4 RDS010 BNC socket to PGZ/PST050 “D/A OUT” socket A95A509(BNC/1.00 m/BNC)5 RDS010 MAB-5M socket to EDI101 MAB-5M socket A95A511(MAB-5M/1.00 m/MAB-5M)6.4.2 SettingsTo control the rotating speed during a sequence run, enter the appropriate valuesfor “D/A OUT initial” and “D/A OUT final” in the method parameters dialogue box. Toobtain a rotating speed of “x” (in rpm) during a method run, enter a “D/A OUT initial”of “x” (in mV). The “D/A OUT final” sets the rotating speed of the EDI101 to get atthe end of the method. Remember the easy rule 1mV=1rpm.Example of a Pot Cyclic Voltammetry:During this voltammetry, the EDI101 will rotate at 1500 rpm. Rotation stops at theend of the method.162VoltaLab 21/06/10/40/50/80Section 6, Chapter 4: Rotating Disc Stand RDS010