Environmental conditionsInterior use onlyAmbient temperature:working range: 5 to 40°Cstorage: -20 to 60°Ctransport: -40 to 60°CRelative humidity:20 to 80 % with temperature between 5 and 31 °C, above 31°C, the interval amplitudedecreases linearly from 20 - 80% at 31°C to 20 - 50% at 40°C.Altitude ≤ 2000 m.Level of pollution : 2Transitory overvoltage: class IIPerformancesStatic features:Output voltage:zero load: ±20 to ±24 VI = ±22 A (HCB020): ±16 to ±18 VI = ±11 A (HCB010): ±16 to ±18 VI = ±5.5 A (HCB005): ±16 to ±18 VAccuracy, voltage compliance: ±0.3%current compliance: ±1.0%voltage measurement: ±0.3%current measurement: ±0.3%Dynamic features (2-electrode setting, WORK electrode to the ground):Slew rate:±10 V; no resistance load < 2 µs (10 to 90 % of the signal)±1 V; ±20 A (R L= 50 mOhm) < 16 µs (10 to 90 % of the signal)Cutout frequency: ±1 V; no resistance load: 200 kHz (-3dB bandwith)±1 V; ±20 A (50 mOhm):25 kHz (-3dB bandwith)Impedance measurements, potentiosatic and galvanostatic modesMeasurement limits (to respect the tolerance of ±3 %, ±3°):< 1 Ohm: 40 kHz0.1 Ohm: 4 kHz0.01 Hz: 400 Hz0.001 Hz: 40 Hz188VoltaLab 21/06/10/40/50/80Section 6, Chapter 8: HCB005 - HCB010 - HCB020 High Current Boosters