6.7.8 Technical specifications - HVB100RegulationMaximum compliance voltage: ±100 VCell ConnectionsREF (CELL) (connection to the cell, Reference Electrode, RE)BNC coaxial socketInput impedance: 10¹² OhmCapacitance < 20 pFAUX (CELL) (connection to the cell, Auxiliary Electrode, AUX)PL259 coaxial socketOutput impedance: LowREF (P) (connection to the PGZ or PST REF socket)BNC coaxial socketAUX (P) (connection to the PGZ or PST AUX socket)PL259 coaxial socketGeneralPower supply120 V~ (±10%) or 230 V~ (±10%) - 50...60 Hz - 170 VAOn/Off power switchSize & Weight485 x 300 x 88 mm & 13 kgMain fusesType : T3.15A L250V (2 fuses)Secondary fuses (not to be replaced by the user)Type : T630mA L250V (2 fuses), T2A L250V (2 fuses)Environmental conditionsInterior use onlyAmbient temperature:working range: 5 to 40°Cstorage: -20 to 60°Ctransport: -40 to 60°CRelative humidity:20 to 80 % with temperature between 5 and 31 °C, above 31°C, the interval amplitudedecreases linearly from 20 - 80% at 31°C to 20 - 50% at 40°C.Altitude ≤ 2000 m.Level of pollution : 2Transitory overvoltage: class II178VoltaLab 21/06/10/40/50/80Section 6, Chapter 7: HVB100 High Voltage Booster