5.2.7 Technical specificationsRegulationCompliance voltage: ±20 VMaximum current output: ±1 APolarization voltage: ±4 VSlew rate: 300 kV/sRise time (100% signal) < 4µsBandwidth (-3dB): 200 kHzCurrent autorangingPotential autorangingVoltammetry (DC)Applied DC potential- Ranges: ±4 V- Best resolution: 125 µV- Accuracy: ±0.2% of rangeMeasured potential- Ranges: ±8 V- Resolution: 0.0125% of range- Best resolution: 1 mV- Accuracy: ±0.2 % of range- Analog output: Yes- DC offset and gain: NoMeasured current- Ranges: 7 from ±1A to ±1µA- Best resolution: 100 pA- Resolution: 0.01% of range- Accuracy: ±0.2 % of range- Analog output: Yes- DC offset and gain: NoFilters- Automatic mode- Manual mode: 5 filters- Anti-oscillation filter: NoScanning performances- Measurement period: 1 s- Max. scan rate: 10 mV/s ( =10 mV/1 s)- Min. scan rate: 125 µV/1 sCell ConnectionsREF (Reference Electrode, RE)- BNC coaxial socket- Input impedance: 10¹² Ohm- Capacitance < 20 pFWORK (Working Electrode, WE)- PL259 coaxial socket- Output impedance: LowAUX (Auxiliary Electrode, CE)- PL259 coaxial socket- Output impedance: LowAdditional channel connectionsVg IN (External Generator)- BNC coaxial socket - Range: ±5 V- Input impedance: 20 kOhmE OUT (Measured potential output)- BNC coaxial socket - Range ±8V- Accuracy: ±0.2%I OUT (Measured current output)- BNC coaxial socket - Range: ± 1V- Accuracy: ±0.2%- Linear with current rangeOther connectionsFunctional ground- banana socket to connect a Faraday cageRS232C (Serial port)- SUB-D9 male socket to connect a PCGeneralPower supply- 120 V~ or 230 V~50...60 Hz - 70 VA- On/Off power switchMains fuses- Type: T800mA L250V (2 fuses)Secondary fuses- Type : T160mA (2 fuses),T400mA (2 fuses) and T1.6A (2 fuses)Size & Weight- 485 x 300 x 88 mm & 6 kgEnvironmental conditions- Interior use only- Ambient temperature:working range: 5 to 40°Cstorage: -20 to 60°Ctransport: -40 to 60°C- Relative humidity: 20 to 80 % withtemperature between 5 and 31 °C, above31°C, the interval amplitude decreaseslinearly from 20 - 80% at 31°C to 20 - 50% at40°C.- Altitude ≤ 2000 m.- Level of pollution : 2- Transitory overvoltage: class IIVoltaLab 21/06/10/40/50/80 147Section 5, Chapter 2: The PGP201 Potentiostat/Galvanostat