6.8.5 Set up the other settingsSet up the other settings parameters as explained in section 3 “Getting started”.For more information, consult the Help file, keyword index:Other settings.6.8.6 Running the experimentSwitch on the PGZ402 and High Current Booster.Switch on the “Cell” button on the High Current Booster, do not switch on the“Cell” button on the PGZ402!Start the experiment from VoltaMaster 4.6.8.7 TroubleshootingThe pilot lamp “ON” is not lit when the High Current Booster is switched onCheck the power supply circuit (mains socket, mains cable, fuse and On/Offswitch). Replacement of the fuses is described below.Replacing the fuses:Switch off the High Current Booster and find the fuse holder on the rear panel.2. Remove the fuse holder1. Remove themains cableFuse rating T10A L250V3. Replace the spent fuse by a new one of the same rating4. Fit the fuse holder back in place (check that the hook is fitted correctly)Warning!Always replace the fuse with the same type and rating.186VoltaLab 21/06/10/40/50/80Section 6, Chapter 8: HCB005 - HCB010 - HCB020 High Current Boosters